header advert
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 136 - 136
1 Jul 2014
Penny J Ding M Ovesen O Varmarken J Overgaard S
Full Access

Summary

Despite high revision rates, the mean two year migration of the ASRTM cup is within an acceptable threshold. Slightly higher migration rates found for the M2a- Magnumâ„¢ Porous Coated Acetabular Component but longer follow up is needed to establish if this implant is at risk.

Introduction

RSA can detect the migration of an implant, and continuous migration is a predictor for failure (1). The ASRTM resurfacing implant was withdrawn from the marked due to excessive failure rate but showed initial femoral component stability. The aim of this study was to investigate the initial implant stability for the ASR cup as a possible explanation for the high revision rate, and to compare it to another metal on metal (MoM) cup.