header advert
Results 1 - 4 of 4
Results per page:
Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 136 - 136
1 Jul 2014
Penny J Ding M Ovesen O Varmarken J Overgaard S
Full Access

Summary

Despite high revision rates, the mean two year migration of the ASRTM cup is within an acceptable threshold. Slightly higher migration rates found for the M2a- Magnum™ Porous Coated Acetabular Component but longer follow up is needed to establish if this implant is at risk.

Introduction

RSA can detect the migration of an implant, and continuous migration is a predictor for failure (1). The ASRTM resurfacing implant was withdrawn from the marked due to excessive failure rate but showed initial femoral component stability. The aim of this study was to investigate the initial implant stability for the ASR cup as a possible explanation for the high revision rate, and to compare it to another metal on metal (MoM) cup.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 94 - 94
1 Sep 2012
Penny J Varmarken J Ovesen O Nielsen C Overgaard S
Full Access

Introduction

Metal on metal articulations produce chromium (Cr) and cobalt (Co) debris, particularly when the articulations are worn in. High levels in the peripheral blood are indicative of excess wear and may cause adverse effects. The present RCT investigates metal ion levels and the relationship of Co, Cr ions and lymphocyte counts during the running-in period.

Materials and Methods

Following randomization to RHA (ASRTM, DePuy) or THA we obtained whole blood (wb), and serum (s) samples at baseline, 8 w, 6 m and 1 y. We measured the Co and Cr concentrations, the total lymphocyte count as well as the CD3+, CD4+, CD8+, CD19+ and CD16+/CD56+ sub populations. Cup inclination and anteversion angles came from conventional radiographs. Activity was measured as steps by pedometer and UCLA activity. Data are presented as median (range).


Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 450 - 451
1 Sep 2009
Penny JO Ovesen O Varmarken J Brixen K Overgaard S
Full Access

Resurfacing THA is claimed to transfer stress naturally to the femur neck and preserve proximal femoral bone mass postoperatively. DXA is an established method in estimating BMD around a standard THA, but due to the anteversion of the femur neck, rotation could affect the size of the neck-regions and thereby the BMD measurements around a RTHA. To our knowledge, this is the first study to analyze the effects of hip rotation on BMD in the femoral neck around a RTHA.

We scanned the femoral neck of 15 patients twice in each position of 15° inward, 0° and 15° outward rotation, and analyzed BMD in a single and a six-region model. CVs were calculated for BMD in the same position as well as between different positions.

For double measurements in the same position we found mean CVs of 3.1% (range 2.5% – 3.7%) and 4.6% (range 2.2% – 8.6%) in the one- and six-region models, respectively. When the 15° outward position was excluded, the CVs decreased to 2.8% and 4.0%. With rotation, the mean CVs rose to 5.4% (range 3.2%–7.2%) and 11.8% (range 2.7% – 36.3%). This effect was most pronounced in the 6-region model, predominantly in the lateral and distal parts of the femoral neck, where the change was significantly different from the fixated position. For the single-region model 15° rotation could be allowed without compromising the precision.

We conclude that rotation adversely affects the precision of BMD measurements around a RTHA, but in the single-region model smaller rotations can be allowed.

With the hip fixated the six-region model produces low CVs, acceptable for longitudinal studies. For maximal topographical detail we prefer the six-region model and recommend that future longitudinal DXA studies, including RTHA, be performed standardised, Preferably, with the hip in the neutral or internal rotation.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 7 | Pages 975 - 979
1 Sep 2003
Therbo M Petersen MM Varmarken J Olsen CA Lund B

Between 1986 and 1991, 106 patients (127 knees) underwent uncemented knee arthroplasty for osteoarthritis. There were 106 total knee arthroplasties and 21 medial unicompartmental knee arthroplasties. The arthroplasties were evaluated for aseptic loosening during the year 2000. For total arthroplasty we used 77 porous-coated anatomic prostheses and 29 press-fit condylar prostheses. The mean bone mineral content of the proximal tibia, measured the day before surgery using dual-photon absorptiometry was 5.48 g/cm for the porous-coated anatomic prostheses which were revised for aseptic loosening (n = 9). This was significantly higher (p = 0.02) than the mean of 4.33 g/cm for those which were not revised. Values for the two revised press-fit condylar knees (4.78 and 4.93 g/cm) were above the mean value (4.23 g/cm) for those which were not revised. We found no statistically significant (p = 0.38) difference between the bone mineral content of the 12 revised and nine unrevised unicompartmental arthroplasties. Low trabecular bone quality, measured as the pre-operative bone mineral content of the proximal tibia, was not a predictor for later revision surgery following uncemented total knee or unicompartmental knee arthroplasty.