Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_4 | Pages 14 - 14
1 Apr 2018
Van Der Straeten C Abdulhussein D Brevadt M Cobb J
Full Access

Background

Hip resurfacing arthroplasty (HRA) and total hip arthroplasty (THA) are treatments of end-stage hip disease. Gait analysis studies comparing HRA and THA have demonstrated HRA results in a more normal gait than THA. The reasons may include the larger, more anatomic head diameter, the preservation of the femoral neck with restoration of the anatomical hip centre position and normal proprioception. This study investigated (1)whether femoral head size diameter affects gait; (2)whether gait still differs between THA and HRA patients even with comparable head diameters.

Methods

We analysed the gait of 33 controls and 50 patients with unilateral hip replacement. Follow-up ranged from 9–68 months. In 27 hips a small femoral head size was used (≤ 36mm); in 23 hips a large head size (>36mm). The small size group consisted of 11 long femoral stem THA and 16 short-stem THA; the large group of 5 long-stem, 8 short-stem THA and 10 HRA patients. There were 14 females/19 males in the control group; 22 females/5 males in the small size group; 13 females/10 males in the large size group.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 90 - 90
1 Apr 2018
Van Der Straeten C Auvinet E Cameron-Blackie A
Full Access

INTRODUCTION

Osteoarthritis (OA) is a growing societal burden, due to the ageing population. Less invasive, less damaging, and cheaper methods for diagnosis are needed, and sound technology is an emerging tool in this field.

AIMS

The aim of the current research was to: 1) investigate the potential of visual scalogram analysis of Acoustic Emission (AE) frequencies within the human audible range (20–20000 Hz) to diagnose knee OA, 2) correlate the qualitative visual scalogram analysis of the AE with OA symptoms, and 3) to do this based on information gathered during gait.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 91 - 91
1 Apr 2018
Chappell K McRobbie D Van Der Straeten C Ristic M Brujic D
Full Access

Purpose

Collagen-rich structures of the knee are prone to damage through acute injury or chronic “wear and tear”. Collagen becomes more disorganised in degenerative tissue e.g. osteoarthritis. An alignment index (AI) used to analyse orientation distribution of collagen-rich structures is presented.

Method

A healthy caprine knee was scanned in a Siemens Verio 3T Scanner. The caprine knee was rotated and scanned in nine directions to the main magnetic field B0. A 3D PD SPACE sequence with isotropic 1×1×1mm voxels (TR1300ms, TE13ms, FOV256mm,) was optimised to allow for a greater angle-sensitive contrast.

For each collagen-rich voxel the orientation vector is computed using Szeverenyi and Bydder's method. Each orientation vector reflects the net effect of all the fibres comprised within a voxel. The assembly of all unit vectors represents the fibre orientation map. Alignment Index (AI) in any direction is defined as a ratio of the fraction of orientations within 20° (solid angle) centred in that direction to the same fraction in a random (flat) case. In addition, AI is normalised in such a way that AI=0 indicates isotropic collagen alignment. Increasing AI values indicate increasingly aligned structures: AI=1 indicates that all collagen fibres are orientated within the cone of 20° centred at the selected direction.

AI = (nM - nRnd)/(nTotal - nRnd) if nM >= nRnd

AI = 0 if nM < nRnd

Where:

nM is a number of reconstructed orientations that are within a cone of 20° centred in selected direction

nRnd is a number of random orientations within a cone of 20° around selected direction

nTotal is a number of collagen reach voxels

By computing AI for a regular gridded orientation space we are able to visualise change of AI on a hemisphere facilitating understanding of the collagen fibre orientation distribution.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 111 - 111
1 Apr 2017
Van Onsem S Lambrecht D Verstraete M Van Der Straeten C Victor J
Full Access

Introduction

Better functional outcomes, lower pain and better stability have been reported with knee designs which restore physiological knee kinematics. Also the ability of the TKA design to properly restore the physiological femoral rollback during knee flexion, has shown to be correlated with better restoration of the flexor/extensor mechanism, which is fundamental to the function of the human knee. The purpose of the study is to compare the kinematics of three different TKA designs, by evaluating knee motion during Activities of Daily Living. The second goal is to see if there is a correlation between the TKA kinematics and the patient reported outcomes.

Methods

Ten patients of each design, who are at least 6 months after their Total Knee Replacement, will be included in this study. Seven satisfied and 3 dissatisfied patients will be selected for each design. In this study 5 different movements will be analysed: flexion/extension; Sitting on and rising from a chair, Stair climbing, descending stairs, Flexion and extension open chain and squatting. These movements will be captured with a fluoroscope. The 2D images that are obtained, will be matched with the 3D implants. This 3D image will be processed with custom-made software to be able to analyse the movement. Tibio-femoral contact points of the medial and lateral condyles, tibio-femoral axial rotation, determination of the pivot-point will be analysed and described. After this analysis, a correlation between the kinematics and the KOOS and KSS will be investigated.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_1 | Pages 5 - 5
1 Jan 2017
Brevadt M Wiik A Aqil A Johal H Van Der Straeten C Cobb J
Full Access

Financial and human cost effectiveness is an increasing evident outcome measure of surgical innovation. Considering the human element, the aim is to restore the individual to their “normal” state by sparing anatomy without compromising implant performance. Gait lab studies have shown differences between different implants at top walking speed, but none to our knowledge have analysed differing total hip replacement patients through the entire range of gait speed and incline to show differences. The purpose of this gait study was to 1) determine if a new short stem femoral implant would return patients back to normal 2) compare its performance to established hip resurfacing and long stem total hip replacement (THR) implants.

110 subjects were tested on an instrumented treadmill (Kistler Gaitway, Amherst, NY), 4 groups (short-stem THR, long-stem THR, hip resurfacing and healthy controls) of 28, 29, 27, and 26 respectively. The new short femoral stem patients (Furlong Evolution, JRI) were taken from the ongoing Evolution Hip trial that have been tested on the treadmill with minimum 12months postop. The long stem total hip replacements and hip resurfacing groups were identified from out 800 patient gait database. They were only chosen if they were 12 months postop and had no other joint disease or medical comorbidities which would affect gait performance.

All subjects were tested through their entire range of gait speeds and incline after having a 5 minute habituation period. Speed intervals were at 0.5kms increments until maximum walking speed achieved and inclines at 4kms for 5, 10, 15%. At all incremental intervals of speed, the vertical component of the ground reaction forces, center of pressure and temporal measurements were collected for both limbs with a sampling frequency of 100Hz. Body weight scaling was applied to correct for mass differences and a symmetry index to compare the implanted hip to the contralateral normal hip. All variables for each subject group were compared to each other using an analysis of variance (ANOVA) with Tukey post hoc test with significance set at α=0.05.

The four experimental groups were reasonably matched for demographics and the implant groups for PROMs. Hip resurfacing had a clear top walking speed advantage, but when assessing the symmetry index on all speeds and incline, all groups were not significantly different. Push-off and step length was statistically less favourable for the short/long THR group (p=0.005–0.05) depending on speed/incline.

The primary aim of this study was determine if implant design affected gait symmetry and performance. Interestingly, irrespective of implant design, symmetry with regards to weight acceptance, impulse, push-off and step length was returned to normal when comparing to healthy controls. However individual implant performance on the flat and incline, showed inferior (p<0.05) push-off force and step length in the short stem and long stem THR groups when compared to controls. Age and gender may have played a part for the short stem group. It appears that the early gait outcomes for the short stem device are promising. Assessment at the 3 year mark should be conclusive.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_2 | Pages 22 - 22
1 Jan 2017
Rivière C Lazennec J Van Der Straeten C Iranpour F Cobb J
Full Access

The current, most popular recommendation for cup orientation, namely the Lewinnek box, dates back to the 70's, that is to say at the stone age of hip arthroplasty. Although Lewinnek's recommendations have been associated with a reduction of dislocation, some complications, either impingement or edge loading related, have not been eliminated. Early dislocations are becoming very rare and most of them probably occur in “outlier” patients with atypical pelvic/hip kinematics. Because singular problems usually need singular treatments, those patients need a more specific personalised planning of the treatment rather than a basic systematic application of Lewinnek recommendations. We aim in this review to define the potential impacts that the spine-hip relations (SHRs) have on hip arthroplasty. We highlight how recent improvements in hip implants technology and knowledge about SHRs can substantially modify the planning of a THR, and make the « Lewinnek recommendations » not relevant anymore. We propose a new classification of the SHRs with specific treatment recommendations for hip arthroplasty whose goal is to help at establishing a personalized planning of a THR. This new classification gives a rationale to optimize the short and long-term patient's outcomes by improving stability and reducing edge loading. We believe this new concept could be beneficial for clinical and research purposes.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_1 | Pages 100 - 100
1 Jan 2017
Navruzov T Rivière C Van Der Straeten C Harris S Cobb J Auvinet E Aframian A Iranpour F
Full Access

The accurate positioning of the total knee arthroplasty affects the survival of the implants(1). Alignment of the femoral component in relation to the native knee is best determined using pre- and post-operative 3D-CT reconstruction(2). Currently, the scans are visualised on separate displays. There is a high inter- and intra-observer variability in measurements of implant rotation and translation(3). Correct alignment is required to allow a direct comparison of the pre- and post-operative surfaces. This is prevented by the presence of the prostheses, the bone shape alteration around the implant, associated metal artefacts, and possibly a segmentation noise.

The aim is to create a novel method to automatically register pre- and post-operative femora for the direct comparison of the implant and the native bone.

The concept is to use post-operative femoral shaft segments free of metal noise and of surgical alteration for alignment with the pre-operative scan. It involves three steps. Firstly, using principal component analysis, the femoral shafts are re-oriented to match the X axis. Secondly, variants of the post-operative scan are created by subtracting 1mm increments from the distal femoral end. Thirdly, an iterative closest point algorithm is applied to align the variants with the pre-operative scan.

For exploratory validation, this algorithm was applied to a mesh representing the distal half of a 3D scanned femur. The mesh of a prosthesis was blended with the femur to create a post-operative model. To simulate a realistic environment, segmentation and metal artefact noise were added. For segmentation noise, each femoral vertex was translated randomly within +−1mm,+−2mm,+−3mm along its normal vector. To create metal artefact random noise was added within 50 mm of the implant points in the planes orthogonal to the shaft. The alignment error was considered as the average distance between corresponding points which are identical in pre- and post-operative femora.

These preliminary results obtained within a simulated environment show that by using only the native parts of the femur, the algorithm was able to automatically register the pre- and post-operative scans even in presence of the implant. Its application will allow visualisation of the scans on the same display for the direct comparison of the perioperative scans.

This method requires further validation with more realistic noise models and with patient data. Future studies will have to determine if correct alignment has any effect on inter- and intra-observer variability.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_2 | Pages 69 - 69
1 Jan 2017
Van Der Straeten C Banica T De Smet A Van Onsem S Sys G
Full Access

Systemic metal ion monitoring (Co;Cr) has proven to be a useful screening tool for implant performance to detect failure at an early stage in metal-on-metal hip arthroplasty. Several clinical studies have reported elevated metal ion levels after total knee arthroplasty (TKA), with fairly high levels associated with rotating hinge knees (RHK) and megaprostheses1. In a knee simulator study, Kretzer2, demonstrated volumetric wear and corrosion of metallic surfaces. However, prospective in vivo data are scarce, resulting in a lack of knowledge of how levels evolve over time. The goal of this study was to measure serum Co and Cr levels in several types TKA patients prospectively, evaluate the evolution in time and investigate whether elevated levels could be used as an indicator for implant failure.

The study was conducted at Ghent University hospital. 130 patients undergoing knee arthroplasty were included in the study, 35 patients were lost due to logistic problems. 95 patients with 124 knee prostheses had received either a TKA (primary or revision) (69 in 55 patients), a unicompartimental knee arthroplasty (7 UKA), a RHK (revision −7 in 6 patients) or a megaprosthesis (malignant bone tumours − 28 in 27 patients). The TKA, UKA and RHK groups were followed prospectively, with serum Co and Cr ions measured preoperatively, at 3,6 and 12 months postoperatively. In patients with a megaprosthesis, metal ions were measured at follow-up (cross-sectional study design).

In primary knees, we did not observe an increase in serum metal ion levels at 3, 6 or 12 months. Two patients with a hip arthroplasty had elevated preTKA Co and Cr levels. There was no difference between unilateral and bilateral knee prostheses. In the revision group, elevated pre-revision levels were found in 2 failures for implant loosening. In both cases, ion levels decreased postoperatively. In revisions with a standard TKA, there was no significant increase in metal ions compared to primary knee arthroplasty. RHK were associated with a significant increase in Co levels even at short-term (3–12 months). The megaprosthesis group had the highest metal ion levels and showed a significant increase in Co and Cr with time in patients followed prospectively. With the current data, we could not demonstrate a correlation between metal ion levels, size of the implant or length of time in situ.

In primary knee arthroplasty with a standard TKA or UKA, metal ion levels were not elevated till one year postoperatively. This suggests a different mechanism of metal ion release in comparison to metal-on-metal hip arthroplasties. In two cases of revision for implant loosening, pre-revision levels were elevated, possibly associated with component wear, and decreased after revision. With RHK, slightly elevated ion levels were found prospectively. Megaprostheses had significantly elevated Co and Cr levels, due to corrosion of large metallic surfaces and/or wear of components which were not perfectly aligned during difficult reconstruction after tumour resection. Further research is needed to assess the clinical relevance of metal ion levels in knee arthroplasty.