Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 84 - 84
1 Sep 2012
Schröder C Utzschneider S Grupp T Fritz B Jansson V
Full Access

Introduction

Minimally invasive implanted unicompartmental knee arthroplasty (UKA) leads to excellent functional results. Due to the reduced intraoperative visibility it is difficult to remove extruded bone cement particles, as well as bone particles generated through the sawing. These loose third body particles are frequently found in minimally invasive implanted UKA.

The aim of this study was to analyse the influence of bone and cement particles on the wear rate of unicompartmental knee prostheses in vitro.

Material & Methods

Fixed- bearing unicompartmental knee prostheses (n = 3; Univation F®, Aesculap, Tuttlingen) were tested with a customized four-station servo-hydraulic knee wear simulator (EndoLab GmbH, Thansau, Germany) reproducing exactly the walking cycle as specified in ISO 14243-1:2002. After 5.0 million cycles crushed cortical bone chips were added to the test fluid for 1.5 million cycles to simulate bone particles, followed by 1.5 million cycles blended with PMMA- particles (concentration of the third-body particles: 5g/l; particle diameter: 0.5- 0.7 mm). Every 500 000 cycles the volumetric wear rate was measured (ISO 14243-2) and the knee kinematics were recorded.

For the interpretation of the test results we considered four different phases: breaking in- (during the first 2.0 million cycles), the steady state- (from 2.0 million to 5 million cycles), bone particle- and cement particle phase.

Finally, a statistical analysis was carried out to verify the normal distribution (Kolmogorov-Smirnov test), followed by direct comparisons to differentiate the volumetric wear amount between the gliding surfaces (paired Student's t-test, p<0.05).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 80 - 80
1 Sep 2012
Utzschneider S Lorber V Dedic M Paulus A Sievers B Jansson V
Full Access

Periprosthetic osteolysis depends on the biological activity of wear particles, but there is little known about the distribution of polyethylene wear particles (PE) in the surrounding joint tissue. The purpose of this study was to examine the localisation of wear particles of six different PEs, including four crosslinked polyethylenes (XPE), as well as their biological activity in the murine knee.

Material and Methods

Wear particles of 4 XPE- and 2 UHMWPE-inserts were isolated (knee joint simulator). For all groups the particles were similar in size and shape (mean diameter 0.3–05μm; 20nm-nucleopore-filter; ISO; n = 100.000).56 female Balb/c mice were randomly assigned to six treatment groups and one control group: control (PBS), XPE1 (3×30 kGy Gamma, annealed/sequential irradiated), XPE2 (95 kGy E-beam, remelted), XPE3 (65 kGy E-beam, remelted), XPE 4 (50 kGy Gamma, remelted), UHMWPE 1, UHMWPE 2. 50 μl of each particle suspension [(0.1% vol/vol (particle volume/PBS volume) after removal of endotoxin] were injected into the left knee joint. After 1 week the mice were killed and a histological and immunhistochemical analysis of the knee joints was done (IL-1, TNF-, ICAM-1). For the immunhistochemistry the articular cartilage, the bone marrow and the synovial membrane were evaluated semiquantitatively (Kruskal-Wallis test; all pairwise multiple comparison procedure; Bonferoni correction; significance level: p<0.05).

Results

All groups showed a thickened synovial layer with an increased cellular infiltration. The particles of XPE 1 and 2 were localised in the bone marrow as well as in the joint space. In contrast, the particles of XPE 3 and 4 were distributed in the synovial layer and in the bone marrow as well, but not in the joint space. The UHMWPE1 particles were mainly located in the bone marrow and joint space while the UHMWPE2 particles were mainly found in the bone marrow and the synovial layer. For all PE groups there was a higher cytokine expression compaired to control (p<0.0024) without any differences between the groups (bone marrow/synovial layer). The chondrocytes in the groups with XPE 1- and XPE 2-particles expressed more TNF- than in the control group and the other treatment groups (p = 0.000).


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 256 - 256
1 Sep 2012
Weber P Schröder C Utzschneider S Jansson V Müller P
Full Access

Introduction

Unicompartmental knee arthroplasty (UKA) in patients with isolated medial osteoarthritis of the knee is nowadays a standard procedure with good results, especially with the minimally-invasive approach. However, the survival rate of the unicompartmental knee prostheses is inferior to that of total knee prostheses. Therefore, further studying of UKA is still necessary. In most mobile bearing designs the femoral component has a spherical surface and therefore its positioning is not crucial. The role of the tibial slope in UKA has not been investigated so far. The manufacturers recommend tibial slopes with values between 10° positive slope and 5° negative slope. Most surgeons try to reconstruct the anatomical slope with a high failure by measuring the slope on x-rays. The aim of this study was to investigate the influence of the tibial slope on the wear rate of a medial UKA.

Materials and methods

In vitro wear simulation of medial mobile bearing unicompartmental knee prosthesis with a spherical femoral surface (Univation ®) was performed with a customized four-station servo-hydraulic knee wear simulator (EndoLab GmbH, Thansau, Germany) reproducing exactly the walking cycle as specified in ISO 14243–1:2002(E). The tibial tray was inserted with 2 different medial tibial slopes: 0°, 8° (n=3 for each group). The lateral tibial slope of the space-holder was not changed (0° for every group). We performed a total of 5 million cycles for every different slope, the gravimetric wear rate was determined gravimetrically using an analytical balance every 500 000 cycles according to the ISO 14243–2.