250 words max Long polished cemented femoral stems, such as the Exeter Hip Revision stem, are one option available to the revision hip arthroplasty surgeon. When proximal bone stock is compromised, distal fixation is often relied upon for stability of the femoral component. In such circumstances, torsional forces can result in debonding and loosening. This study compared the torsional behaviour of a cemented polished and featureless (plain) stem with cemented, polished stems featuring fins or flutes. Nine torsional tests were carried out on each of these three different stem designs. The finned stem construct was significantly stiffer than the fluted stem (mean 24.5 Nm/deg v 17.5 Nm/deg). The plain stem mean stiffness was less than the featured stems (13 Nm/deg), but wide variability lead to no statistically significant difference. The maximum torque of the finned (30.5 Nm) and fluted stems (29 Nm) was significantly higher than the plain stem (10.5 Nm); with no significance to the difference between the finned and fluted stems. Distal stem features may provide a more reliable and greater resistance to torque in polished, cemented revision hip stems. Finned stem features may also increase the stiffness of the construct. Consideration should thus be given to the incorporation of distal stem features in the design of revision hip stems.