Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 29 - 29
1 Apr 2017
Verstraete M Van Onsem S Stevens C Herregodts S Arnout N Victor J
Full Access

Background

In-vitro testing of knee joints remains vital in the understanding of knee surgery and arthroplasty. However, based on the design philosophy of the original Oxford knee simulator, this in-vitro testing has mainly focused on squatting motion. As the activities of daily living might drastically differ from this type of motion, both from a kinematic and kinetic point of view, a new knee simulator is required that allows studying more random motion patterns. This paper describes a novel knee simulator that overcomes the limitations of traditional Oxford simulators, providing both kinematic and kinetic freedom with respect to the applied boundary conditions.

Methods

This novel test simulator keeps the hip at a fixed position, only providing a single rotational degree of freedom (DOF) in the sagittal plane. In addition, the ankle holds four DOF, including all rotational DOF and the translation along the medio-lateral axis. Combining these boundary conditions leaves five independent DOF to the knee; the knee flexion angle is actively controlled through the positioning of the ankle joint in the antero-posterior and proximal-distal direction. The specimens' quadriceps muscle is actively controlled, the medial and lateral hamstrings are passively loaded. To validate the performance of this simulator, two fresh frozen specimens have been tested during normal squatting and cycling. Their kinematic patterns have been compared to relevant literature data.