When performing revision total hip arthroplasty using diaphyseal-engaging titanium tapered stems (TTS), the recommended 3 to 4 cm of stem-cortical diaphyseal contact may not be available. In challenging cases such as these with only 2 cm of contact, can sufficient axial stability be achieved and what is the benefit of a prophylactic cable? This study sought to determine, first, whether a prophylactic cable allows for sufficient axial stability when the contact length is 2 cm, and second, if differing TTS taper angles (2° vs 3.5°) impact these results. A biomechanical matched-pair cadaveric study was designed using six matched pairs of human fresh cadaveric femora prepared so that 2 cm of diaphyseal bone engaged with 2° (right femora) or 3.5° (left femora) TTS. Before impaction, three matched pairs received a single 100 lb-tensioned prophylactic beaded cable; the remaining three matched pairs received no cable adjuncts. Specimens underwent stepwise axial loading to 2600 N or until failure, defined as stem subsidence > 5 mm.Aims
Methods
Periprosthetic femur fracture (PFF) is a major complication following total hip arthroplasty (THA) that carries significant morbidity, mortality, and economic burden. Currently, uncemented stems are highly preferred in primary THA, but have been associated with higher risk of PFF compared to cemented stems. The use of collared stems in uncemented primary THA has shown promise in reducing PFF rates postoperatively. This retrospective study included 2,294 uncemented primary THAs using the posterior approach performed by two attending surgeons from January 2016 to December 2022. Both surgeons switched from a collarless femoral stem design to a collared design in May 2020. Data was collected regarding stem design, frequency of PFF, and requirement for revision surgery. Periprosthetic fractures were identified and confirmed using medical records and/or radiographic imaging. Fracture rates and percentages between collared and collarless stems were then analyzed. A Fisher's Exact Test was performed to determine if there was a significant association between collared and collarless stem use on PFF rates. A total of 2,294 uncemented primary THAs performed by 2 surgeons were eligible for analysis. 903 (39.4%) patients received a collared stem, and 1,391 (60.6%) patients received a collarless stem. In total, 14 (0.6%) PFFs occurred over the study period. There was 1 fracture (0.1%) out of 903 collared stems, and 13 fractures (0.9%) out of 1,391 collarless stems (p = 0.012). Collared stems were associated with a significant decrease in PFF rate when compared to collarless stems in uncemented primary THA. Future studies are encouraged to continue to investigate PFF and other complication rates with the use of a collared stem design.
Although CT is considered the benchmark to measure femoral version, 3D biplanar radiography (hipEOS) has recently emerged as a possible alternative with reduced exposure to ionizing radiation and shorter examination time. The aim of our study was to evaluate femoral stem version in postoperative total hip arthroplasty (THA) patients and compare the accuracy of hipEOS to CT. We hypothesize that there will be no significant difference in calculated femoral stem version measurements between the two imaging methods. In this study, 45 patients who underwent THA between February 2016 and February 2020 and had both a postoperative CT and EOS scan were included for evaluation. A fellowship-trained musculoskeletal radiologist and radiological technician measured femoral version for CT and 3D EOS, respectively. Comparison of values for each imaging modality were assessed for statistical significance.Aims
Methods
Navigation devices are designed to improve a surgeon’s accuracy in positioning the acetabular and femoral components in total hip arthroplasty (THA). The purpose of this study was to both evaluate the accuracy of an optical computer-assisted surgery (CAS) navigation system and determine whether preoperative spinopelvic mobility (categorized as hypermobile, normal, or stiff) increased the risk of acetabular component placement error. A total of 356 patients undergoing primary THA were prospectively enrolled from November 2016 to March 2018. Clinically relevant error using the CAS system was defined as a difference of > 5° between CAS and 3D radiological reconstruction measurements for acetabular component inclination and anteversion. Univariate and multiple logistic regression analyses were conducted to determine whether hypermobile (Aims
Methods
Spinopelvic mobility plays an important role in functional acetabular component position following total hip arthroplasty (THA). The primary aim of this study was to determine if spinopelvic hypermobility persists or resolves following THA. Our second aim was to identify patient demographic or radiological factors associated with hypermobility and resolution of hypermobility after THA. This study investigated patients with preoperative posterior hypermobility, defined as a change in sacral slope (SS) from standing to sitting (ΔSSstand-sit) ≥ 30°. Radiological spinopelvic parameters, including SS, pelvic incidence (PI), lumbar lordosis (LL), PI-LL mismatch, anterior pelvic plane tilt (APPt), and spinopelvic tilt (SPT), were measured on preoperative imaging, and at
six weeks and a minimum of
one year postoperatively. The severity of bilateral hip osteoarthritis (OA) was graded using Kellgren-Lawrence criteria.Aims
Methods
The anatomic dual mobility (ADM) technology utilized a monoblock cobalt chromium acetabular component. However, design limitations conferred difficulties controlling orientation during component insertion and inability to confirm full implant seating; the solution resulted in the creation of the modular dual mobility (MDM). The modular implant combines a standard titanium acetabular component and a cobalt chromium liner insert. Due to the metal-on-metal interface on MDM implants, fretting and corrosion releasing metal ions like previous metal-on-metal THA implants, were a concern. This study prospectively reviewed metal ions (cobalt, chromium and titanium) on patients who were at least 1 year post MDM implantation and compared them to patients with an ADM implant and evaluated radiographic seating of the components. All patients with ADM and MDM implants underwent evaluation of metal ions (cobalt, chromium and titanium) at their one year follow-up appointment. Radiographic evaluation for acetabular polar gaps was performed. Elevated metal ions were determined using standard laboratory ranges. Differences in baseline demographics were assessed using the Mann Whitney-U test and Fishers's exact test. Differences in metal ions and implant type were compared using the Fisher's exact tests.Introduction
Methods
Custom flange acetabular components (CFACs) are a patient-specific option for addressing large acetabular defects at revision total hip arthroplasty (THA), but patient and implant characteristics that affect survivorship remain unknown. This study aimed to identify patient and design factors related to survivorship. A retrospective review of 91 patients who underwent revision THA using 96 CFACs was undertaken, comparing features between radiologically failed and successful cases. Patient characteristics (demographic, clinical, and radiological) and implant features (design characteristics and intraoperative features) were collected. There were 74 women and 22 men; their mean age was 62 years (31 to 85). The mean follow-up was 24.9 months (Aims
Patients and Methods
Instability continues to be a troublesome complication after total hip arthroplasty (THA). Patient-related risk factors associated with a higher dislocation risk include the preoperative diagnosis, an age of 75 years or older, high body mass index (BMI), a history of alcohol abuse, and neurodegenerative diseases. The goal of this study was to assess the dislocation rate, radiographic outcomes, and complications of patients stratified as high-risk for dislocation who received a dual mobility (DM) bearing in a primary THA at a minimum follow-up of two years. We performed a retrospective review of a consecutive series of DM THA performed between 2010 and 2014 at our institution (Hospital for Special Surgery, New York, New York) by a single, high-volume orthopaedic surgeon employing a single prosthesis design (Anatomic Dual Mobility (ADM) Stryker, Mahwah, New Jersey). Patient medical records and radiographs were reviewed to confirm the type of implant used, to identify any preoperative risk factors for dislocation, and any complications. Radiographic analysis was performed to assess for signs of osteolysis or remodelling of the acetabulum.Aims
Materials and Methods
Patients with rheumatoid arthritis (RA) have a higher risk of surgical site infection (PJI) than patients with osteoarthritis (OA). Disease modifying therapy is in widespread use in RA patients, and biologic medications may increase Staphylococcus aureus colonization rates. Because S. aureus colonization likely increases risk of surgical infection, perioperative assessments and therapies to decrease the risk of invasive S.aureus infections may be warranted. The objective of this study was to determine if there was a difference in S. aureus carriage among patients with RA, OA, and RA on biologics (RA+B). An a priori power analysis determined 123 participants per group were needed to detect a relative difference of 20% among groups with 80% power. After IRB approval, patients were screened; included patients met American College of Rheumatology classification criteria. Patients were approached between April 2017 and May 2018 and asked to perform a nasal swab while on site using the Center for Disease Control's swabbing protocol; questionnaires pertaining to their current health status were collected. Swabs were inoculated onto ChromAgar/ChromID MRSA plates for detection of S. aureus. Mann-Whitney U and Chi-square tests were used to evaluate baseline differences between groups. Logistic regression evaluated the associations between groups and S. aureus carriage. All statistical analyses were performed using SAS Software version 9.3 (SAS Institute, Cary, NC); statistical significance was defined as p<0.05.Introduction
Methods
We report on the outcome of the Synergy cementless femoral stem
with a minimum follow-up of 15 years (15 to 17). A retrospective review was undertaken of a consecutive series
of 112 routine primary cementless total hip arthroplasties (THAs)
in 102 patients (112 hips). There were 60 female and 42 male patients
with a mean age of 61 years (18 to 82) at the time of surgery. A
total of 78 hips in the 69 patients remain Aims
Patients and Methods
The aim of this systematic review was to report the rate of dislocation
following the use of dual mobility (DM) acetabular components in
primary and revision total hip arthroplasty (THA). A systematic review of the literature according to the Preferred
Reporting Items for Systematic Reviews and Meta-analyses guidelines
was performed. A comprehensive search of Pubmed/Medline, Cochrane
Library and Embase (Scopus) was conducted for English articles between
January 1974 and March 2016 using various combinations of the keywords “dual
mobility”, “dual-mobility”, “tripolar”, “double-mobility”, “double
mobility”, “hip”, “cup”, “socket”. The following data were extracted
by two investigators independently: demographics, whether the operation
was a primary or revision THA, length of follow-up, the design of
the components, diameter of the femoral head, and type of fixation
of the acetabular component.Aims
Materials and Methods
We aimed to quantify the relative contributions of the medial
femoral circumflex artery (MFCA) and lateral femoral circumflex
artery (LFCA) to the arterial supply of the head and neck of the
femur. We acquired ten cadaveric pelvises. In each of these, one hip
was randomly assigned as experimental and the other as a matched
control. The MFCA and LFCA were cannulated bilaterally. The hips
were designated LFCA-experimental or MFCA-experimental and underwent
quantitative MRI using a 2 mm slice thickness before and after injection
of MRI-contrast diluted 3:1 with saline (15 ml Gd-DTPA) into either
the LFCA or MFCA. The contralateral control hips had 15 ml of contrast
solution injected into the root of each artery. Next, the MFCA and
LFCA were injected with a mixture of polyurethane and barium sulfate
(33%) and their extra-and intra-arterial course identified by CT
imaging and dissection.Aims
Materials and Methods
This study investigates and defines the topographic
anatomy of the medial femoral circumflex artery (MFCA) terminal
branches supplying the femoral head (FH). Gross dissection of 14
fresh–frozen cadaveric hips was undertaken to determine the extra
and intracapsular course of the MFCA’s terminal branches. A constant
branch arising from the transverse MFCA (inferior retinacular artery;
IRA) penetrates the capsule at the level of the anteroinferior neck,
then courses obliquely within the fibrous prolongation of the capsule
wall (inferior retinacula of Weitbrecht), elevated from the neck,
to the posteroinferior femoral head–neck junction. This vessel has
a mean of five (three to nine) terminal branches, of which the majority
penetrate posteriorly. Branches from the ascending MFCA entered
the femoral capsular attachment posteriorly, running deep to the
synovium, through the neck, and terminating in two branches. The
deep MFCA penetrates the posterosuperior femoral capsular. Once
intracapsular, it divides into a mean of six (four to nine) terminal
branches running deep to the synovium, within the superior retinacula
of Weitbrecht of which 80% are posterior. Our study defines the
exact anatomical location of the vessels, arising from the MFCA
and supplying the FH. The IRA is in an elevated position from the
femoral neck and may be protected from injury during fracture of
the femoral neck. We present vascular ‘danger zones’ that may help
avoid iatrogenic vascular injury during surgical interventions about
the hip. Cite this article: