Element analysis was performed by EDX (Oxford D. 7060) to identify carbides and the alloy composition. Element distribution maps were taken to separate the single elements.
All implants showed a sphericity deviation less than 10μm. On average the heads tended to have a higher spherical deviation of 4.1μm (SD: 2.3μm) compared to the cups 2.7μm (SD: 1.4μm). Based on the SEM and EDX inspection the manufacturing process, heat treatment and carbide distribution could be clarified.
The characteristically unspheric formations of the heads may be due to the cooling process after manufacturing the implant and there is also a relation between the wall thickness of the implant and the unspheric formations. With decreasing wall thickness the implant cools faster locally. Additionally a cup with a thin wall may deform under loading condition and a very tight clearance could be detrimental. This study will help to understand clinical observations. It still has to be proven that these biomechanical factors influence the clinical performance of hip resurfacing implants.
A fifth control group did not receive any implant. Animals were followed for 3 months after transplantation and than sacrificed. The histological specimens were evaluated by image analysis after immuno-histochemical stains were performed (including the following antigens – collagen type II, collagen type I, collagen type III, collagen type X, S-100, alkaline phosphatase, osteocalcin, osteopontin, nitric oxide synthase).