header advert
Results 1 - 1 of 1
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXII | Pages 16 - 16
1 Jul 2012
Granville-Chapman J Hacker A Keightley A Sarkhel T Monk J Gupta R
Full Access

Extensor tendon ruptures have been reported in up to 8.8% of patients after volar plating and long screws have been implicated. The dihedral dorsal surface of the distal radius hinders accurate screw length determination using standard radiographic views (lateral; pronation and supination). A ‘dorsal tangential’ view has recently been described, but has not been validated.

To validate this view, we mounted a plate-instrumented sawbone onto a jig. Radiographs at different angles were reviewed independently by 11 individuals. Skyline views clearly demonstrated all screw tips, whereas only 69% of screw tips were identifiable on standard views.

With screws 2mm proud of the dorsal surface, skyline views detected 67% of long screws (sensitivity). The best of the standard views achieved only 11% sensitivity. At 4mm long, skyline sensitivity was 85%, compared with 25% for standard views. At 6mm long, 100% of long screws were detected on skylines, but only 50% of 8mm long screws were detected by standard views. Inter and intra-observer variability was 0.97 (p=0.005).

For dorsal screw length determination of the distal radius, the skyline view is superior to standard views. It is simple to perform and its introduction should reduce the incidence of volar plate-related extensor tendon rupture.