Vol. 105-B, Issue SUPP_7 | Pages 139 - 139
4 Apr 2023
Mühling M Sandriesser S Augat P
Full Access

Implant manufacturers develop new products to improve existing fracture fixation methods or to approach new fracture challenges. New implants are commonly tested and approved with respect to their corresponding predecessor products, because the knowledge about the internal forces and moments acting on implants in the human body is unclear. The aim of this study was to evaluate and validate implant internal forces and moments of a complex physiological loading case and translate this to a standard medical device approval test.

A finite elements model for a transverse femur shaft fracture (AO/OTA type 32-B2) treated with a locked plate system (AxSOS 3 Ti Waisted Compression Plate Broad, Stryker, Kalamazoo, USA) was developed and experimentally validated. The fractured construct was physiologically loaded by resulting forces on the hip joint from previously measured in-vivo loading experiments (Bergmann et. al). The forces were reduced to a level where the material response in the construct remained linear elastic. Resulting forces, moments and stresses in the implant of the fractured model were analysed and compared to the manufacturers’ approval data.

The FE-model accurately predicted the behaviour of the whole construct and the micro motion of the working length of the osteosynthesis. The resulting moment reaction in the working length was 24 Nm at a load of 400 N on the hip. The maximum principle strains on the locking plate were predicted well and did not exceed 1 %.

In this study we presented a protocol by the example of locked plated femur shaft fracture to calculate and validate implant internal loading using finite element analysis of a complex loading. This might be a first step to move the basis of development of new implants from experience from previous products to calculation of mechanical behaviour of the implants and therefore, promote further optimization of the implants’ design.