Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_2 | Pages 20 - 20
1 Feb 2020
Mueller J Bischoff J Siggelkow E Parduhn C Roach B Drury N Bandi M
Full Access

Introduction

Initial stability of cementless total knee arthroplasty (TKA) tibial trays is necessary to facilitate biological fixation. Previous experimental and computational studies describe a dynamic loading micromotion test used to evaluate the initial stability of a design. Experimental tests were focused on cruciate retaining (CR) designs and walking gait loading. A FEA computational study of various constraints and activities found CR designs during walking gait experienced the greatest micromotion. This experimental study is a continuation of testing performed on CR and walking gait to include a PS design and stair descent activity.

Methods

The previously described experimental method employed robotic loading informed by a custom computational model of the knee. Different TKA designs were virtually implanted into a specimen specific model of the knee. Activities were simulated using in-vivo loading profiles from instrumented tibia implants. The calculated loads on the tibia were applied in a robotic test. Anatomically designed cementless tibia components were implanted into a bone surrogate. Micromotion of the tray relative to the bone was measured using digital image correlation at 10 locations around the tray.

Three PS and three CR samples were dynamically loaded with their respective femur components with force and moment profiles simulating walking gait and stair descent activities. Periods of walking and stair descent cycles were alternated for a total of 2500 walking cycles and 180 stair descent cycles. Micromotion data was collected intermittently throughout the test and the overall 3D motion during a particular cycle calculated. The data was normalized to the maximum micromotion value measured throughout the test. The experimental data was evaluated against previously reported computational finite element model of the micromotion test.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 55 - 55
1 Apr 2019
Mueller JK Roach B Parduhn C
Full Access

Introduction

Cementless total knee arthroplasty (TKA) designs are clinically successful and allow for long term biological fixation. Utilizing morselized bone to promote biological fixation is a strategy in cementless implantation. However, it is unknown how bone debris influences the initial placement of the tray. Recent findings show that unseated tibia trays without good contact with the tibial resection experience increased motion. This current study focuses on the effect of technique and instrument design on the initial implantation of a cementless porous tibia. Specifically, can technique or instrument design influence generation of bone debris, and thereby change the forces required to fully seat a cementless tray with pegs?

Methods

This bench top test measured the force-displacement curve during controlled insertion of a modern cementless tibia plate with two fixation pegs. A total of nine pairs of stripped human cadaver tibias were prepared according to the surgical technique. However, the holes for the fixation pegs were drilled intentionally shallow to isolate changes in insertion force due to the hole preparation. A first generation instrument set (Instrument 1.0) and new instrument set design (Instrument 2.0), including a new drill bit designed to remove debris from the peg hole, were used. The tibias prepared with Instrument 1.0 were either cleaned to remove bone debris from the holes or not cleaned. The tibias prepared with the Instrument 2.0 instruments were not cleaned, resulting in three groups: Instrument 1.0 (n=7), Instrument 1.0 Cleaned (n=5), and Instrument 2.0 (n=6). Following tibia resection and preparation of holes for the fixation pegs, the tibias were cut and potted in bone cement ensuring the osteotomy was horizontal. The tibial tray was mounted in a load frame (Enduratec) and the trays were inserted at a constant rate (0.169mm/sec) while recording the force. The test was concluded when the pegs were clearly past the bottom of the intentionally shallow holes.