Current advice regarding implant choice is based on estimates of cost-benefit derived from implant survival to an endpoint of revision. Current estimates do not account for many implant failures which are treated with non-revision surgery and may not be accurate. The aim of this study was to estimate survival of major stem implant design groups to an endpoint of reoperation.

Primary total hip replacement and linked revision form the National Joint Registry (NJR) and Hospital Episode Statistics (HES) data linked by unique identifier were used. Survival of femoral implant groups (cemented stainless steel polished taper [PTSS], cemented cobalt chrome polished taper [PTCC], cemented composite beam [CB], collarless cementless [NCOL] and collared cementless [COL]) was estimated using Kaplan-Meier method.

809,832 patients with valid NJR and HES data from England, were included. Cumulative failure at ten years for PTSS increased overall from 2.9% (95%CI 2.8–2.9) to 3.6% (95%CI 3.6–3.7) after inclusion of reoperations. Cumulative failure at ten years for PTSS increased from 2.5% (95%CI 2.5–2.6) to 3.3% (95%CI 3.2–3.4), for PTCC increased from 3.8% (95%CI 3.5–4.0) to 5.4% (95%CI 5.1–5.6), for CB increased from 3.1% (95%CI 2.9–3.3) to 4.1% (95%CI 3.8–4.3), for NCOL increased from 3.4% (95%CI 3.3–3.5) to 3.9% (95%CI 3.8–4.0), and for COL increased from 2.5% (95%CI 2.4–2.6) to 3.1% (95%CI 2.9–3.2), after inclusion of reoperations.

Re-operation for internal fixation is as significant life event for the patient as revision. When a more inclusive metric is used, the patient and clinician's perspective on what constitutes a GIRFT implant may not be the same. Further work is required to update implant selection guidance in view of the change in implant performance.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_6 | Pages 64 - 64
2 May 2024
Lamb J West R Relton S Wilkinson M Pandit H
Full Access

Current estimates of periprosthetic fracture risk associated with femoral implants is mostly limited to revision only estimates and does not accurately represent stem performance. The aim of this study was to estimate the risk of surgically treated post-operative periprosthetic femoral fracture (POPFF) more accurately associated with frequently used femoral implants used for total hip arthroplasty (THA).

A cohort study of patients who underwent primary THA in England between January 1, 2004, and December 31, 2020. Periprosthetic fractures were identified from prospectively collected revision records and national procedure coding records. Survival modelling was used to estimate POPFF incidence rates, adjusting for potential confounders. Subgroup analyses were performed for patients over 70 years, with non-osteoarthritic indications, and neck of femur fracture.

POPFF occurred in 0.6% (5100/809,832) of cases during a median (IQR) follow up of 6.5 (3.9 to 9.6) years. The majority of POPFF were treated with fixation after implantation of a cemented stem. Adjusted patient time incidence rates (PTIR) for POPFF varied by stem design, regardless of cement fixation. Cemented composite beam stems (CB stems) demonstrated the lowest risk of POPFF. Collared cementless stems had an equivalent or lower rate of POPFF versus the current gold standard polished taper slip cemented stem.

POPFF account for a quarter of all revisions following primary THA. Cemented CB stems are associated with the lowest POPFF risk. Stem design is strongly associated with POPFF risk, regardless of the presence of cement. Surgeons, policymakers, and patients should consider these findings when recommending femoral implants in those most at risk of POPFF.