We carried out 60 revision procedures for failed porous coated anatomic total knee replacements in 54 patients, which were divided into two groups. The 14 knees in group I had a well-fixed femoral component at surgery which was retained, and in the 46 knees in group II both tibial and femoral components were loose and were revised using a variety of implants. Our review comprised clinical and radiological assessment. A total of 13 knees required a second revision. Six (42%) in group I failed very early (mean 2.1 years) when compared with seven (15%) in group II (mean 6.8 years). Failure was due to wear of the polyethylene insert by the abraded, retained femoral component (crude odds ratio 4.07; 95% CI 1.07 to 15.5). We recommend a complete change of primary bearing surfaces at the time of revision of an uncemented total knee replacement in order to prevent early wear of polyethylene.
Method: 66 revision procedures for failed uncemented porous coated anatomic (PCA) total knee replacement were performed in 60 patients. At review, four patients had died while two were lost to follow up, therefore 60 knees in 54 patients were included in this prospective study. The principal indications for primary revision were polyethylene wear and loosening of the tibial base plate. 14 patients had a well fixed femoral component and hence were retained while 46 patients had both the components revised. All patients were prospectively assessed prior to surgery, at three months, six months and yearly thereafter. Review comprised clinical and radiological assessment. The mean follow up was 8.4 yrs (7–12 yrs)