Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 154 - 154
1 Dec 2013
Raja LK Yanoso-Scholl L Nevelos J Schmidig G Thakore M
Full Access

Introduction

Frictional torque is generated at the hip joint during normal gait loading and motion [1]. This study investigated the effect of shell deformation due to press-fit on frictional torque generated at the articulating surfaces of cementless acetabular shells that incorporated fixed and dual mobility bearing designs.

Materials and Methods

Figure 1 lists the study groups (minimum of n = 5). All groups were tested with a 50 mm Trident PSL shell (Stryker Orthopaedics, NJ) and a Ti6Al4V trunnion. Metal-on-Metal specimens were custom designed and manufactured, and are not approved for clinical use. The remaining groups consisted of commercially available products (Stryker Orthopaedics, NJ).

All groups were tested with the shells in deformed and undeformed states.

Deformed Setup: A two-point relief configuration was created in a polyurethane foam block (Figure 2) with a density of 30 lb/ft3 to replicate shell deformation due to press-fit [2]. The blocks were machined to replicate the press-fit prescribed in the shell's surgical protocol. Each shell was assembled into the foam block by applying an axial force at 5 mm/min until it was completely seated.

Undeformed Setup: Each shell was assembled in a stainless steel block with a hemispherical cavity that resulted in a line-to-line fit with the shell OD.

Frictional torque was measured using a physiologically relevant test model [3]. In this model, the specimen block was placed in a fixture to simulate 50° abduction and 130° neck angle (Figure 2). A 2450N side load was applied and the femoral head underwent angular displacement of ± 20° for 100 cycles at 0.75 Hz. The articulating surfaces were lubricated with 25% Alpha Calf Fraction Serum.

Peak torque was observed towards the end or the beginning of each cycle where the velocity of the femoral head approaches 0 and the head changes direction. This torque is referred as maximum static frictional torque. Specimen groups were statistically compared with a single-factor ANOVA test and a Tukey post-hoc test at 95% confidence level. Paired t-tests were performed to compare individual groups in deformed and undeformed states.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 489 - 489
1 Dec 2013
Yanoso-Scholl L Raja LK Nevelos J Longaray J Herrera L Schmidig G Thakore M
Full Access

Introduction

Many tests have been published which measure frictional torque [1–4] in THR. However, different test procedures were used in those studies. The purpose of this study was to determine the effect of test setup on the measured friction torque values.

Methods

Specimen Description Table 1 lists tested study groups (n≥3). Metal-on-Metal specimens were custom designed and manufactured, and are not approved for clinical use. The remaining groups consisted of commercially available products (Stryker Orthopaedics, NJ).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 490 - 490
1 Dec 2013
Yanoso-Scholl L Raja LK Schmidig G Heffernan C Thakore M Nevelos J
Full Access

Introduction

The femoral head/stem taper modular junction has several advantages; it also has the potential to result in fretting [1]. Stability of the taper junction is critical in reducing the risk associated with fretting. The purpose of this test was to measure the strength of various commercially available head-stem taper combinations under torsional loads to determine the effect of taper geometry and material on the strength of this taper junction.

Methods and Materials

CoCr femoral heads were tested with trunnions that were machined with both a large and small taper geometry, replicating commercially available stem taper designs, V40 (small) and C (large) (Table-1, Stryker Orthopaedics, NJ).

The femoral heads were assembled onto the trunnions with a 2 kN axial force. A multi-axis test frame (MTS Corp, MN) was used to test the head-trunnion combination by dynamically loading with a torque of ± 5Nm and a constant axial load of 2450N for 1000 cycles at 1.5 Hz (Figure 1). Samples were submerged in 25% diluted Alpha Calf Fraction Serum (Hyclone, UT). Upon completion of the dynamic test, a static torque to failure test was performed where the axial force of 2450N was maintained and the trunnion was rotated to 40° at a rate of 3°/sec.

The torque required to rotate the trunnion by 1° was determined for each specimen. Also, the torsional resistance, defined as change in torque/change in angle in the linear region of the torque-angular displacement data curve, was calculated for all the specimens. A limitation associated with the static test was that at 1° rotation it was difficult to differentiate between rotation of the trunnion inside the femoral head and physical twisting of the trunnion. Specimen groups were compared with a single-factor ANOVA test and a Tukey post hoc test at 95% confidence level.