Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:

Introduction

Despite the multiple patho-aetiological basis of Hallux valgus (HV), corrective osteotomy is a common and globally performed orthopaedic procedure. Over-correction of the inter-metatarsal angle (IMA), however, is not without consequence. Through increased joint pressures, over-correction may predispose to joint degeneration. Hypothesizing that over-correction leads to an increase in intra-articular joint pressures, we constructed a mechanical simulation of the 1st metatarsal-phalangeal joint.

Methods

The vector forces of Flexor Hallucis Longus, Extensor Hallucis Longus, Extensor Hallucis Brevis and Adductor Hallucis, acting upon the 1st proximal phalanx were simulated with tensioned weighted rope. A Foley catheter balloon inflated to 1.5ml, simulated intra-articular distractive vector forces. The joint was freely mobile.

At 5° IMA intervals from 10° valgus to 15° varus, intrarticular pressures were recorded using a contact pressure (Graseby™, Hertfordshire, UK) transducer attached to a calibrated manometer (Digitech Instruments™, Ulhasnagar, India). At each angle, measurements were repeated five times with removal and recalibration of the pressure monitor after each measurement.