Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 304 - 304
1 Dec 2013
Arno S Fetto J Bell C Papadopoulos K Walker P
Full Access

INTRODUCTION:

The purpose of this study was to determine if a short femoral stem (Lima Corporate, Udine, Italy) would result in a strain distribution which mimicked the intact bone better than a traditional length stem, thereby eliminating the potential for stress-shielding.

METHODS:

A 2 mm thick moldable plastic (PL-1, Vishay Micromeasurements, Raleigh, NC) was contoured to six fourth-generation composite femoral bones (Pacific Research Laboratories, Vashon, WA). The intact femurs were then loaded (82 kg) in a rig which simulated mid-stance single limb support phase of gait (Figure 1). During testing, the femurs were viewed and video recorded through a model 031 reflection polariscope. Observing the photoelastic coating through the polariscope, a series of fringes could be seen, which represented the difference in principal strain along the femur. The fringes were quantified using Fringe Order, N, as per the manufacturers technical notes. In order to analyze the strain distribution, the femur was separated into 6 zones, 3 lateral and 3 medial, and the maximum fringe order determined. Upon completion of testing of the intact femur, the short length femoral stem was inserted and tested, and finally the traditional length femoral stem was inserted and tested. Anterior and lateral radiographs were obtained of the femur with each femoral stem in order to confirm proper alignment.