Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 84 - 84
1 Sep 2012
Pamu J Hussain A Daniel J Kamali A Li C
Full Access

Introduction

Ion analysis has been used as one of the key indicators to assess the performance of MoM devices in patients. Modular devices, in particular having larger overall surface area (the stem and sleeve), and locking interfaces (head – bore, sleeve- taper and sleeve-bore, stem-taper surfaces) than other MoM devices are expected to release greater number of ions. Concerns have been expressed that the ion release at the taper junction might be a potential cause leading to the failure of the implant [Garbuz et al, 2010].

The aim of this study was to look into the wear and the associated ion release from the taper junction and the articulating surface of modular devices.

Method

For the first time a novel design has been used to isolate the taper junction on modular devices on the hip simulators in order to compare the wear at the taper junction and articulating surface. The taper junction has been isolated in a small gaiter, while the head and cup were contained in a large gaiter. CoCrMo sleeves having an offset of +8 mm have been used on 50 mm modular heads along with Ti6Al4V stems. The acetabular components were standard BHR cups. Three devices (Smith & Nephew, UK) have been tested with newborn calf serum as a lubricant (in the large gaiter) and also as the medium containing the taper junction (in the small gaiter). The serum samples from the articulating surface and taper junction were analysed using HR-ICPMS. The locking interfaces at the taper junction have been left intact throughout the duration of the test. Both the head and the cup have been tested under anatomical conditions using the standard implant development centre's (IDC) profile for 2 million cycles (Mc). The lubricant was newborn calf serum with 0.2% sodium azide diluted with de-ionised water to achieve protein concentration of 20 mg/ml. The flexion/extension was 30°/15° and the internal/external rotation was ±10°. The force was Paul-type stance phase loading with a maximum load of 3 kN and a standard ISO swing phase load of 0.3 kN. The frequency was 1 Hz, with an 8 hour stop after every 16 hours of testing.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 71 - 71
1 Sep 2012
Hussain A Kamali A Li C Pamu J
Full Access

Metal-on-Metal devices generate significantly lower volumetric wear than conventional total hip replacements. However, clinically some patients may suffer some form of laxity in their joints leading to subluxation of the joint, which in turn may cause edge loading of an implant thereby increasing the chances of failure due to higher than expected wear.

In this study, the effect of subluxation on MoM implant wear was investigated on a hip joint simulator.

Materials & Methods

Two groups of 44 mm MoM devices were tested, n=3 in each group. The devices were subjected to 1 and 2 mm of subluxation. The flexion/extension was 30° and 15° respectively, internal/external rotation was ±10°, and cup inclination was 35°. The force was Paul type stance phase loading with a maximum load of 3 kN, with ISO swing phase load of 0.3 kN, run at 1 Hz.

The test was carried out on a ProSim deep flexion & subluxation hip wear simulator (SimSol, UK). Rather than separating the head and the cup (microseparation), or reducing the swing phase load, this simulator is equipped with a novel mechanism to achieve translation of the head, while subjecting the devices to subluxation. During the swing phase, a controlled lateral force necessary for the translation of the head is applied by a cam mechanism, head retraction will then take place on heel strike.

The lubricant used was new born calf serum with 0.2 wt. % sodium azide concentration diluted with de-ionised water to achieve average protein concentration of 20 g/l. Lubricant was changed every 250k cycles. Gravimetric wear measurements have been taken at 0.25 & 0.5 Mc stages.

Results

Tests conducted with 1mm (Group 1) and 2mm (Group 2) subluxation significantly increased volumetric wear compared to standard hip simulator tests [1]. At 0.5 million cycles, group 1 and 2 produced an average volume loss of 4.38±0.98 mm3 (95% CL) and 7.07±1.64 mm3 (95% CL) respectively.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXV | Pages 108 - 108
1 Jun 2012
Pamu J Kamali A Daniel J Hussain A Li C
Full Access

Introduction

Hip implant research has been carried out for decades using hip simulators to reflect situations in vivo. With regards to metal on metal (MoM) implant testing, it has been reported that there is no significant difference between the wear generated by various cobalt chromium (CoCr) microstructures. On the contrary, higher wear, metal ion levels and subsequent failures have been reported in heat treated (high carbon, low carbide) devices compared to as cast (high carbon, high carbide) devices in vivo. During testing, the bearing surfaces may be masked from the effect of microstructure on wear under fast and continuous cycles, while in vivo, the extensive range of kinetics and kinematics, stop/start motion, varying walking frequencies could break down the fluid film, resulting in a less favourable lubrication regime. The aims of this study were to develop a more physiologically relevant hip simulator test protocol, and investigate the effect of microstructure on wear.

Materials & Methods

Three pairs of 50mm as cast (AC) and four pairs of 50mm double heat treated (DHT) CoCr MoM devices were tested in a ProSim hip simulator. In order to determine the frequency for testing, Patients' activities have been monitored using a Step Activity Monitor (SAM) device. The data showed a relatively slower walking pace (frequency) than that used in the hip simulator studies. The new frequency, along with stop/start motion and various kinetics and kinematics profiles have been used in putting together a more physiologically relevant hip simulator test protocol. The lubricant used in this study was new born calf serum with 0.2 % (w/v) sodium azide concentration diluted with de-ionised water to achieve an average protein concentration of 20 g/l. Gravimetric measurements have been taken at 0.5, 1, 1.5 & 2 million cycle (Mc) stages and ion analysis has been carried out on the serum samples.