Introduction

Iatrogenic proximal femur hoop-stress fracture is a recognised complication of uncemented hip arthroplasty. It has a reported incidence of two to three percent and increases patient morbidity. We describe a novel technology that predicts fracture in real-time by less than one minute.

Method

Four proximal femora from red deer (Cervus elaphus), similar size to human proximal femora, were prepared to accept an uncemented hip arthroplasty femoral rasp (Finsbury Orthopaedics) and then mounted in a loading machine. The femora were fresh-frozen, defrosted and kept at room temperature in 0.9% saline swabs. The rasp was forced into each femur in repeated loading cycles every 10 seconds, in steps of 100N increasing from 200N to over 2000N until fracture, in a manner to simulate surgery. One sensor was attached to the surface of the proximal femur and one to the femoral rasp. The sensor outputs were recorded, analysed and displayed on a PC using a software algorithm to show signal energy (joules) and amplitude (decibels). The proximal femur was coated with specular marking paint to permit real-time 3-D digital image correlation (DIC) analysis. DIC is an established tool in engineering fracture analysis and utilises two spatially orientated video cameras to measure surface strain and fracture. The femur was observed by the human eye and loaded in cycles until a fracture was seen. The moment of fracture was marked in the recording timeline. DIC was used to confirm fracture.