Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 34 - 34
1 Apr 2019
Bandi M Oswald A Siggelkow E
Full Access

Introduction

In knee biomechanics the concept of the envelope of motion (EOM) has proven to be a powerful method to characterize joint mechanics and the effect of surgical interventions. It is furthermore indispensable for numerical model validation. While commonly used for tibiofemoral kinematics, there is very little report of applying the concept to patellofemoral kinematics. EOM measurements require precise and reproducible displacement and load control in all degrees of freedom (DOF), which robotic testing has proven to provide. The objectives of this study were therefore to (1) develop a robotic method to assess patellofemoral EOM as a function of tibiofemoral EOM, (2) compare resulting patellofemoral kinematics to published data, and (3) determine which DOFs in the tibiofemoral EOM mostly account for the patellofemoral EOM.

Material and Methods

The developed robotic (KUKA KR140 comp) method was evaluated using 8 post-mortem human leg specimens of both genders (age: 55±11 years, BMI: 23±5). Firstly, tibiofemoral neutral flexion was established as well as the EOM by applying anterior-posterior (±100 N), medial-lateral (±100 N), internal-external (±4 Nm) and varus-valgus (±12 Nm) loads under low compression (44 N) at 7 flexion angles. Secondly, patellofemoral flexion kinematics and EOM were measured during a robotic playback of the previously established tibiofemoral kinematics. During these measurements, the quadriceps tendon was loaded with a hanging weight (20 kg) via a pulley system directing the force to the anterior superior iliac spine. Kinematics were tracked optically (OptiTrack) and registered to CT scans using co-scanned aluminum cylinders and beads embedded in the patella.

The overall patellofemoral EOM was calculated as the extent of patellar motion observed during manipulating the tibia inside the tibiofemoral EOM in all DOFs. Additionally, patellofemoral EOMs were calculated for tibial manipulations along individual DOFs to analyze the importance of these DOFs.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 33 - 33
1 Apr 2019
Bandi M Siggelkow E Oswald A Parratte S Benazzo F
Full Access

Introduction

Partial knee arthroplasty (PKA) has demonstrated the potential to improve patient satisfaction over total knee arthroplasty. It is however perceived as a more challenging procedure that requires precise adaptation to the complex mechanics of the knee. A recently developed PKA system aims to address these challenges by anatomical, compartment specific shapes and fine-tuned mechanical instrumentation. We investigated how closely this PKA system replicates the balance and kinematics of the intact knee.

Materials and Methods

Eight post-mortem human knee specimens (age: 55±11 years, BMI: 23±5, 4 male, 4 female) underwent full leg CT scanning and comprehensive robotic (KUKA KR140 comp) assessments of tibiofemoral and patellofemoral kinematics. Specimens were tested in the intact state and after fixed bearing medial PKA. Implantations were performed by two experienced surgeons.

Assessments included laxity testing (anterior-posterior: ±100 N, medial-lateral: ±100 N, internal-external: ±3 Nm, varus- valgus: ±12 Nm) under 2 compressive loads (44 N, 500 N) at 7 flexion angles and simulations of level walking, lunge and stair descent based on in-vivo loading profiles. Kinematics were tracked robotically and optically (OptiTrack) and represented by the femoral flexion facet center (FFC) motions. Similarity between intact and operated curves was expressed by the root mean square of deviations (RMSD) along the curves. Group data were summarized by average and standard deviation and compared using the paired Student's T-test (α = 0.05).