Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 130 - 130
1 Jul 2014
Schneider K Zderic I Gueorguiev B Richards R Nork S
Full Access

Summary

Biomechanically, a 2° screw deviation from the nominal axis in the PFLCP leads to significantly earlier implant failure. Screw deviation relies on a technical error on insertion, but in our opinion cannot be controlled intraoperatively with the existing instrumentation devices.

Background

Several cases of clinical failure have been reported for the Proximal Femoral Locking Compression Plate (PFLCP). The current study was designed to investigate the failure mode and to explore biomechanically the underlying mechanism. Specifically, the study sought to determine if the observed failure was due to technical error on insertion or due to implant design.