In 1988 we reported a ten-year review of 83 surviving patients from a group of 135 (146 prostheses) who had undergone primary hip replacement using the Stanmore prosthesis. We have now reviewed 44 of these patients at 15 to 16 years. Four patients had undergone revision, but the other 40 were all satisfied with the result of their hip replacement, 36 having little or no pain. Functional activities had decreased, but were still adequate for their average age of 81 years. There had been definite migration of the cup and/or femoral component in three hips, wear of the cup in ten and resorption of the calcar in six. Of the 24 hips inserted with radiopaque cement, eight showed an increase in radiolucent lines at the acetabular interface. The cumulative survival rate of the prosthesis was 91% at 15 to 16 years.
Between February 1975 and August 1976, 195 total hip replacements using Stanmore components were performed; of these, 146 were in 135 patients who had not had previous hip surgery. At review 52 had died, but none of the others was lost to follow-up. Of the 52, two had had a second operation, one for infection and one for recurrent dislocation. In the remaining 83 patients (92 hips) five revisions were necessary: four for aseptic loosening, and one for stem fracture. The remaining 78 patients had little or no pain and little restriction of activity. After a follow-up period of at least nine years, the survival rate of the prosthesis was 95%. There had been migration of the femoral component in five cases and migration of the acetabular cup in one case, but no wear of the acetabular component could be demonstrated.