Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_5 | Pages 55 - 55
1 Apr 2019
Mueller JK Roach B Parduhn C
Full Access

Introduction

Cementless total knee arthroplasty (TKA) designs are clinically successful and allow for long term biological fixation. Utilizing morselized bone to promote biological fixation is a strategy in cementless implantation. However, it is unknown how bone debris influences the initial placement of the tray. Recent findings show that unseated tibia trays without good contact with the tibial resection experience increased motion. This current study focuses on the effect of technique and instrument design on the initial implantation of a cementless porous tibia. Specifically, can technique or instrument design influence generation of bone debris, and thereby change the forces required to fully seat a cementless tray with pegs?

Methods

This bench top test measured the force-displacement curve during controlled insertion of a modern cementless tibia plate with two fixation pegs. A total of nine pairs of stripped human cadaver tibias were prepared according to the surgical technique. However, the holes for the fixation pegs were drilled intentionally shallow to isolate changes in insertion force due to the hole preparation. A first generation instrument set (Instrument 1.0) and new instrument set design (Instrument 2.0), including a new drill bit designed to remove debris from the peg hole, were used. The tibias prepared with Instrument 1.0 were either cleaned to remove bone debris from the holes or not cleaned. The tibias prepared with the Instrument 2.0 instruments were not cleaned, resulting in three groups: Instrument 1.0 (n=7), Instrument 1.0 Cleaned (n=5), and Instrument 2.0 (n=6). Following tibia resection and preparation of holes for the fixation pegs, the tibias were cut and potted in bone cement ensuring the osteotomy was horizontal. The tibial tray was mounted in a load frame (Enduratec) and the trays were inserted at a constant rate (0.169mm/sec) while recording the force. The test was concluded when the pegs were clearly past the bottom of the intentionally shallow holes.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 109 - 109
1 Sep 2012
Mueller JK Sharma A Komistek R Meccia B
Full Access

Orthopaedic companies spend years and millions of dollars developing and verifying new total knee arthroplasty (TKA) designs. Recently, computational models have been used in the hopes of increasing the efficiency of the design process. The most popular predictive models simulate a cadaveric rig. Simulations of these rigs, although useful, do not predict in vivo behavior. Therefore, in this current study, the development of a physiological forward solution, or predictive, rigid body model of the knee is described.

The models simulate a non-weight bearing extension activity or a weight-bearing deep knee bend (DKB) activity. They solve for both joint forces and kinematics simultaneously and were developed from the ground up. The models are rigid body and use Kane's dynamical equations. The model began with a simple two dimensional non-weight bearing extension activity model of the tibiofemoral joint. Step by step the model was expanded. Quadriceps and hamstring muscles were added to drive the motion. Ligaments were added represented by multiple non-linear spring elements. The model was expanded to three-dimensions (3D) allowing out of plane motions and calculation of medial and lateral condylar forces. The patella was added as its own body allowing for simulation of the patellofemoral joint. The model was then converted to a weight bearing deep knee bend activity. A pelvis and trunk were added and muscles were given physiological origin and insertion points. A modified proportional-integral-derivative (PID) controller was implemented to control the rate of flexion and also to assist in joint stability by adjusting the force in individual quadriceps muscles. A method for representing articulating geometry was developed. Once the deep knee bend model was fully developed (Figure 1) it was converted back to a non-weight bearing extension model (Figure 2) resulting in simulations of a normal knee performing a weight bearing and non-weight bearing activity. The tibiofemoral kinematic results were compared to in vivo kinematics obtained from a fluoroscopy study of five normal subjects. Parameters from the CT models of one of these subjects (Subject 3) were used in the model.

The model kinematics behave as the normal knee does in vivo. The kinetic results were within reasonable ranges with a maximum total quadriceps force of 0.86 BW and 4.73 BW for extension and DKB simulations, respectively (Figure 3 and Figure 4). The maximum total tibiofemoral forces were 1.26 BW and 3.70 BW for extension and DKB, respectively. The relationship between the quadriceps force, patella ligament force and patellofemoral forces are consistent with how the extensor mechanism behaves (Figure 3 and Figure 4). The patellofemoral forces are low between 0 and 20 degrees flexion and the patella ligament and quadriceps forces are close in magnitude from 0 to around 70 degrees flexion when the patellofemoral forces increase and the quadriceps forces increase relative to the patella ligament force. The model allows for virtual implantation of TKA geometry and after kinematic and kinetic validation from in vivo TKA data can be used to predict the behavior of TKA in vivo.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XL | Pages 225 - 225
1 Sep 2012
Zingde S Leszko F Mueller JK Mahfouz M Dennis D Komistek R
Full Access

INTRODUCTION

Posterior stabilized (PS) total knee arthroplasty (TKA) provides posterior stability with the use of a cam-post mechanism which performs the function of the posterior cruciate ligament. The tibial post engages with the femoral cam, prevents the femur from sliding anteriorly and provides the posterior femoral rollback necessary for achieving deep flexion of the knee. However, these designs do not substitute the resection of the anterior cruciate ligament. In order to overcome this deficit, other TKA designs have been recently introduced to provide dual support, with the help of dual cam-post engagement mechanism. Various studies conducted on the PS TKA have suggested that the cam-post mechanism does not engage as designed, resulting in tibial post wear and increased stresses resulting in backside wear of the polyethylene insert component. Also, the in vivo data pertaining to the actual cam-post engagement mechanism in bi-cruciate stabilized knees is still very limited. Therefore, the objective of this study was to determine the cam-post mechanism interaction under in vivo, weight bearing conditions for subjects implanted with either a Rotating Platform (RP) Posterior Stabilized (PS) TKA or a bi-cruciate stabilizing TKA (BCS).

METHODS

In-vivo, weight-bearing, 3D knee kinematics were determined for eight subjects (9 knees) having a RP-PS TKA (DePuy Inc.) and eight subjects (10 knees) having BCS TKA (Smith&Nephew Inc.), while performing a deep knee bend. 3D kinematics was recreated from the fluoroscopic images using a previously published 3D-to-2D registration technique (Figure 1). Images from full extension to maximum flexion were analyzed at 10° intervals. Once the 3D kinematics of all implant components was recreated, the cam-post mechanism was scrutinized. The distance between the interacting surfaces was monitored throughout the flexion and the predicted contact map was calculated. The instances, when the minimum distance between the cam and post surfaces dropped to zero was considered to indicate the engagement of the mechanism. This analysis was carried out for both the, anterior and posterior cam-post engagement sites.