Bone & Joint Research
Vol. 7, Issue 2 | Pages 196 - 204
1 Feb 2018
Krull A Morlock MM Bishop NE

Objectives

Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of this study was to determine the effects of increasing the impaction energy and the stiffness of the impactor tool on the stability of the taper junction and on the forces transmitted through the patient’s surrounding tissues.

Methods

A commercially available impaction tool was modified to assemble components in the laboratory using impactor tips with varying stiffness at different applied energy levels. Springs were mounted below the modular components to represent the patient. The pull-off force of the head from the stem was measured to assess stability, and the displacement of the springs was measured to assess the force transmitted to the patient’s tissues.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 218 - 218
1 Jul 2014
Huber G Nagel K Skrzypiec DM Klein A Püschel K Morlock MM
Full Access

Summary

Lumbar spinal specimens exhibited high fatigue strength. The cycles to failure are not only dependent on the maximum peak load, but also on the load offset or the amplitude, respectably.

Introduction

Spinal injury might be caused by whole body vibrations. The permitted exposure to vibration in the workplace is therefore limited. However, there is a lack in knowledge how external vibrations might cause internal damages. Numerical whole body models might provide the potential to estimate the dynamic spinal loading during different daily activities, but depends on knowledge about the corresponding fatigue strength. This study is aiming to determine the in vitro fatigue strength of spinal specimens from donors of working age.