Periprosthetic fracture (PF) after primary total hip arthroplasty
(THA) is an uncommon but potentially devastating complication. This
study aims to investigate the influence of cemented stem designs
on the risk of needing a revision for a PF. We analysed data on 257 202 primary THAs with cemented stems
and 390 linked first revisions for PF recorded in the National Joint
Registry (NJR) of England, Wales and Northern Ireland to determine
if a cemented femoral stem brand was associated with the risk of
having revision for a PF after primary THA. All cemented femoral
stem brands with more than 10 000 primary operations recorded in
the NJR were identified. The four most commonly used cemented femoral
stems were the Exeter V40 (n = 146 409), CPT (n = 24 300), C-Stem
(n = 15 113) and Charnley (n = 20 182). We compared the revision risk ratios due to PF amongst the stems
using a Poisson regression model adjusting for patient factors.
Compared with the Exeter V40, the age, gender and ASA grade adjusted
revision rate ratio was 3.89 for the cemented CPT stem (95% confidence
interval (CI) 3.07 to 4.93), 0.89 for the C-Stem (95% CI 0.57 to
1.41) and 0.41 for the Charnley stem (95% CI 0.24 to 0.70).Aims
Patients and Methods
This study aimed to: (1) compare published follow-up guidelines for metal-on-metal (MoM) hip patients and analyse protocols in relation to current evidence, and (2) assess the financial implications of these guidelines. Follow-up guidance for MoM hips from five national authorities (MHRA in the UK; EFORT; United States FDA; Therapeutic Goods Administration of Australia; Health Canada) were contrasted and critically appraised. Using National Joint Registry (NJR) data (67,363 MoM hips implanted) the cost of annual surveillance for all MoM hips recorded in the NJR was calculated for each protocol.Introduction
Methods