High tibial osteotomy (HTO) is an effective surgical treatment for isolated medial compartment knee osteoarthritis; however, widespread adoption is limited due to difficulty in achieving the planned correction, and patient dissatisfaction due to soft tissue irritation. A new HTO system – Tailored Osteotomy Knee Alignment (TOKA®, 3D Metal Printing Ltd, Bath, UK) could potentially address these barriers having a custom titanium plate and titanium surgical guides featuring a unique mechanism for precise osteotomy opening as well as saw cutting and drilling guides. The aim of this study was to assess the accuracy of this novel HTO system using cadaveric specimens; a preclinical testing stage ahead of first-in-human surgery according to the ‘IDEAL-D’ framework for device innovation. Local ethics committee approval was obtained. The novel opening wedge HTO procedure was performed on eight cadaver leg specimens. Whole lower limb CT scans pre- and post-operatively provided geometrical assessment quantifying the discrepancy between pre-planned and post-operative measurements for key variables: the gap opening angle and the patient specific surgical instrumentation positioning and rotation - assessed using the implanted plate. The average discrepancy between the pre-operative plan and the post-operative osteotomy correction angle was: 0.0 ± 0.2°. The R2 value for the regression correlation was 0.95. The average error in implant positioning was −0.4 ± 4.3 mm, −2.6 ± 3.4 mm and 3.1 ± 1.7° vertically, horizontally, and rotationally respectively. This novel HTO surgery has greater accuracy and smaller variability in correction angle achieved compared to that reported for conventional or other patient specific methods with published data available. This system could potentially improve the accuracy and reliability of osteotomy correction angles achieved surgically.