The primary aim of this independent prospective randomised trial
was to compare serum metal ion levels for ceramic-on-metal (CoM)
and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty
(THA). Our one-year results demonstrated elevation in metal ion
levels above baseline with no significant difference between the
CoM and MoM groups. This paper reviews the five-year data. The implants used in each patient differed only in respect to
the type of femoral head (ceramic or metal). At five-year follow-up
of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available
for comparison.Aims
Patients and Methods
In a double-blinded randomised controlled trial,
83 patients with primary osteoarthritis of the hip received either
a ceramic-on-metal (CoM) or metal-on-metal (MoM) total hip replacement
(THR). The implants differed only in the bearing surfaces used.
The serum levels of cobalt and chromium and functional outcome scores
were compared pre-operatively and at six and 12 months post-operatively. Data were available for 41 CoM and 36 MoM THRs (four patients
were lost to follow-up, two received incorrect implants). The baseline
characteristics of both cohorts were similar. Femoral head size
measured 36 mm in all but two patients who had 28 mm heads. The
mean serum cobalt and chromium levels increased in both groups,
with no difference noted between groups at six months (cobalt p
= 0.67, chromium p = 0.87) and 12 months (cobalt p = 0.76, chromium
p = 0.76) post-operatively. Similarly, the mean Oxford hip scores,
Western Ontario and McMaster Universities Osteoarthritis index and
University of California, Los Angeles activity scores showed comparable improvement
at 12 months. Our findings indicate that CoM and MoM couplings are associated
with an equivalent increase in serum cobalt and chromium levels,
and comparable functional outcome scores at six and 12-months follow-up.
We report the effectiveness of revision of total
elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows
in 52 patients were treated with re-cementing of a total elbow replacement
into part or all of the existing cement mantle or into the debrided
host-bone interface, without the use of structural bone augmentation
or a custom prosthesis. The original implant revision was still
This study assesses long term performance of the Souter-Strathclyde elbow arthroplasty. From 1984 to 1999, 68 Souter-Strathclyde prostheses were inserted in 53 patients; nine patients died, one was lost and 38 (88%) had full clinical examination. Mean survival was 72 months (range 8 to 187), 25 elbows survived to 5-year follow up, with improvement in pain, motion, stability and function. Mayo score was satisfactory in 92% of all 68 elbows. Complications occurred in 13 elbows (19%) and 14 elbows were revised for instability (six), fracture (three), loosening (three) and intraoperative problems (two). Survivorship at 10 years was 74% (95% Confidence Interval ± 7. 7).