Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_1 | Pages 146 - 146
1 Feb 2020
King C Chakour K Kim Y Luu H Martell J
Full Access

Introduction

Background: Trochanteric bursitis is a common and poorly understood complication following total hip arthroplasty (THA). The purpose of this study was to evaluate the incidence of symptomatic trochanteric bursitis and the change in hip offset among THA patients before and after the introduction of robotic assistance.

Methods

Retrospective chart review of THAs performed by a single over a 3-year period between 1/5/2013 and 6/28/2016. Between 1/5/2013 and 11/11/2014 101 consecutive patients were identified that underwent manual posterior-lateral THA that utilized traditional cup positioning method based on AP Pelvis radiograph. The subsequent six-month period during a complete transition to robotic arm assistance for posterior-lateral THA was excluded to eliminate any learning curve or selection bias. Between 6/2015 and 6/2016 109 consecutive patients that underwent robotic arm-assisted. Medical records were reviewed for symptomatic trochanteric bursitis within two years of surgery. Hip offset was measured on preoperative and postoperative AP pelvis radiographs and postoperative joint reactive forces were calculated using Martell's Hip Analysis Suite.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 108 - 108
1 Mar 2013
Higgs G Kurtz S Hanzlik J MacDonald D Kane WM Day J Klein GR Parvizi J Mont M Kraay M Martell J Gilbert J Rimnac C
Full Access

Introduction

Wear debris generation in metal-on-metal (MOM) total hip arthroplasty (THA) has emerged as a compelling issue. In the UK, clinically significant fretting corrosion was reported at head-taper junctions of MOM hip prostheses from a single manufacturer (Langton 2011). This study characterizes the prevalence of fretting and corrosion at various modular interfaces in retrieved MOM THA systems used in the United States.

Methods and Materials

106 MOM bearing systems were collected between 2003 and 2012 in an NIH-supported, multi-institutional retrieval program. From this collection, 88 modular MOM THA devices were identified, yielding 76 heads and 31 stems (22 modular necks) of 7 different bearing designs (5 manufacturers) for analysis. 10 modular CoCr acetabular liners and 5 corresponding acetabular shells were also examined. Mean age at implantation was 58 years (range, 30–85 years) and implantation time averaged 2.2 ± 1.8 years (range, 0–11.0 years). The predominant revision reason was loosening (n=52). Explants were cleaned and scored at the head taper, stem taper, proximal and distal neck tapers (for modular necks), liner, and shell interfaces in accordance with the semi-quantitative method of Goldberg et al. (2002).