Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 72 - 72
1 Apr 2018
Santos I Mahmoud M Thorwächter C Bourgeois A Müller P Pietschmann M Chevalier Y
Full Access

Background

While total shoulder arthroplasty (TSA) is a generally successful procedure, glenoid loosening remains a common complication. Though the occurrence of loosening was related to patient-specific factors, biomechanical factors related to implant features may also affect the fixation of the glenoid component, in particular increased glenohumeral mismatch that could result in eccentric loads and translations. In this study, a novel test setup was used to quantify glenohumeral pressures for different motion patterns after TSA.

Methods

Six cadaveric human shoulders were implanted with total shoulder replacements (Exactech, Inc., USA) and subjected to cyclic internal-external, flexion-extension and abduction-adduction rotations in a passive motion testing apparatus. The system was coupled to a pressure sensor system (Tekscan, Inc., USA) to acquire joint loads and to a Zebris system (Zebris Medical, GmbH, Germany) to measure joint kinematics. The specimens were subjected to a total of 2160 cycles and peak pressures were compared for each motion pattern.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 16 - 16
1 Apr 2018
Klar R Bondarava M Wiedenmann T Seitz D Ren B Thasler W Müller P Betz O
Full Access

Since the development of biomimetic and ceramic bone reconstructive in the early 1970, these specialised bioreactors intended for bone or cartilage regeneration have come a long way in trying to design an alternative procedure other than autogenous bone grafting. However, all known biomaterials still fall short of inducing substantial bone formation in vitro or in vivo, especially when treating large bony defects. As such there is a necessity to develop novel bone-reconstructive biomaterials that can more appropriately be utilised and can induce substantial more bone formation than current scaffolds. Using the rapid prototyping technique (Friedrich-Baur BioMed Center, Bayreuth, Germany) to develop new and improved hydroxyapatite/β-tricalcium phosphate devices, which can be predesigned to any outer shape with controlled pore structure and exhibit a unique intrinsic porosity <150µm due to the 3D-printing process to fit any skeletal bone loss site, the aim of our laboratories was to test the osteoinductive capacity of these new bioreactors in an in vitro culture system utilising adipose-derived stem cells (ADSCs). Immunofluorescent staining revealed that beside the standard surface protein expression patterns typical for ADSCs, the cells also produced osteoblast specific proteins, specifically osteocalcin, osteopontin and dentin matrix acidic phosphoprotein 1. ADSCs seeded on the surface of the biomimetic scaffolds showed constant proliferation, maintained viability and differentiation throughout the scaffold, including the small intrinsic pores. Subsequent, qRT-PCR also revealed that alkaline phosphatase and osteocalcin expression was significantly increased upon addition of osteogenic medium but even more so when human recombinant morphogenetic protein 2 (hBMP-2) was included. Immunofluorescent data of protein expression was consistent with qRT-PCR data. Taken into account with previous results by our laboratories in respect to adipose tissue as a viable inductive medium that can form substantial new bone formation in vivo the present results demonstrated that the investigated bioceramic devices possess the necessary capacity that could, together with adipose tissue, provide the next leap necessary to finally and decisively induce substantial or total new bone formation in clinical bone defects of humans.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 75 - 75
1 Apr 2018
Matsuura M Schmidutz F Sprecher C Müller P Chevalier Y
Full Access

Introduction

Stemless shoulder implants have recently gained increasing popularity. Advantages include an anatomic reconstruction of the humerus with preservation of bone stock for upcoming revisions. Several implant designs have been introduced over the last years. However, only few studies evaluated the impact of the varying designs on the load transfer and bone remodeling. The aim of this study was to compare the differences between two stemless shoulder implant designs using the micro finite element (µFE) method.

Materials and Methods

Two cadaveric human humeri (low and high bone mineral density) were scanned with a resolution of 82µm by high resolution peripheral quantitative computer tomography (HR-pQCT). Images were processed to allow virtual implantation of two types of reverse-engineered stemless humeral implants (Implant 1: Eclipse, Arthrex, with fenestrated cage screw and Implant 2: Simpliciti, Tornier, with three fins). The resulting images were converted to µFE models consisting of up to 78 million hexahedral elements with isotropic elastic properties based on the literature. These models were subjected to two loading conditions (medial and along the central implant axis) and solved for internal stresses with a parallel solver (parFE, ETH Zurich) on a Linux Cluster. The bone tissue stresses were analysed according to four subregions (dividing plane: sagittal and frontal) at two depths starting from the bone-implant surface and the distal region ending distally from the tip of Implant 1 (proximal, distal)


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 74 - 74
1 Apr 2018
Chevalier Y Chamseddine M Santos I Müller P Pietschmann M
Full Access

Introduction

Glenoid loosening, still a main complication for shoulder arthroplasty, was suggested to be related implant design, surgical aspects, and also bone quality. However, typical studies of fixation do not account for heterogeneity in bone morphology and density which were suggested to affect fixation failure. In this study, a combination of cyclic rocking horse tests on cadaver specimens and microCT-based finite element (microFE) analysis of specimens of a wide range of bone density were used to evaluate the effects of periprosthetic bone quality on the risks of loosening of anatomical keeled or pegged glenoid implants.

Methods

Six pairs of cadaveric scapulae, scanned with a quantitative computer tomography (QCT) scanner to calculate bone mineral density (BMD), were implanted with either cemented anatomical pegged or keeled glenoid components and tested under constant glenohumeral load while a humeral head component was moved cyclically in the inferior and superior directions. Edge displacements were measured after 1000, 4000 and 23000 test cycles, and tested for statistical differences with regards to changes or implant design. Relationships were established between edge displacements and QCT-based BMD below the implant. Four other specimens were scanned with high-resolution peripheral QCT (82µm) and implanted with the same 2 implants to generate virtual models. These were loaded with constant glenohumeral force, varying glenohumeral conformity and superior or inferior load shifts while internal stresses at the cement-bone and implant-cement interfaces were calculated and related to apparent bone density in the periprosthetic zone.