Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Applied filters
Hip

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_12 | Pages 32 - 32
1 Jun 2017
Di Laura A Hothi H Henckel J Liow M Kwon Y Skinner J Hart A
Full Access

Dual mobility (DM) cups are designed to improve stability, however have been associated with increased risk of impingement that can ultimately result in intraprosthetic dislocation. It is speculated that the femoral neck plays a role in their performance. We investigated the effect of neck topography on the wear of new-generation liners.

This was a retrieval study involving 70 DM cups implanted with liners made of highly crosslinked polyethylene and paired with two neck types: either highly polished (n=35) or rough necks (n=35). The median time of implantation was 30 months.

The rim edge of all inserts was investigated by two examiners for evidence of contact with the femoral neck, presenting as deformation of the polyethylene. A high precision roundness machine and micro-CT scans of the components were used to measure the size of the deformations observed.

28 of the 35 (80%) DM liners paired with rougher necks had evidence of neck impingement resulting in a raised lip, whilst 8 out of 35 (23%) liners paired with smooth necks had a raised lip; this difference was significant (p<0.0001). The repeatability and the inter-observer reproducibility of the deformation scores was found to be substantial κ >0.70. The height of the raised rims of the DM cups paired with rough necks had a median (range) of 139 µm (72–255), whilst had a median (range) of 52 µm (45–90) with smooth necks, the difference between the groups was significant (p<0.0001).

Liner rim deformation resulting from contact with the femoral neck likely begins during early in-vivo function.

Rough necks can increase the damage on the polyethylene rim in dual-mobility bearing, which may lead to loss of the retentive power of these components over time.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_12 | Pages 51 - 51
1 Jun 2017
Di Laura A Hothi H Henckel J Liow M Kwon Y Skinner J Hart A
Full Access

Dual-taper implants provide surgeons with options to optimise patients' anatomy intraoperatively but are at risk of early revision due to adverse tissue reactions to corrosion debris. Risk factors for failure and linkage with symptoms however are not fully understood. We related retrieval findings to clinical and implant variables.

This study involved 88 failed dual-taper implants with TMZF femoral stems and cobalt-chromium necks, revised for pain, elevated Co (median = 7.3μg/L) and Cr (median = 2.15μg/L) ions levels and fluid collection on MRI.

Stem-neck surfaces were assessed for: 1) severity of corrosion using a published visual method and 2) severity of material lost and location of damage with a roundness-measuring machine. Five traces were taken on each round section of the taper surface at 45° increments to compute the relative depth of damage. The total area of these traces provided a measure of surface damage for comparative purposes.

The stem-neck taper junctions were severely corroded; the deepest areas of damage were on the inferior-proximal and superior-distal part of the necks, compatible with cantilever bending. Elemental analysis revealed chromium rich deposits indicative of corrosion processes and metal transfer from the stem to the neck.

There was a positive correlation between the severity of damage and time of implantation (p<0.0001). Co and Cr levels in the blood were also strongly correlated (p<0.0001, p=0.0002). No other implant or patient variables were linked.

The stem-neck junction was severely corroded in all cases. The severity and location of the areas of surface damage did not link with implant or patient characteristics in this big cohort suggesting that the design and material combination is the predominant source of failure in these designs.

Dual-taper hips are severely corroded at the stem-neck junction; this appears to be due to the use of a TMZF alloy stem paired with CoCrMo necks.