Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 27 - 27
1 Mar 2009
Lautridou C Lebel B Burdin G Leclercq S Vielpeau C
Full Access

Background: Early or later dislocation is a frequent complication associated with total hip arthroplasty. The Bousquet’s acetabular component, an original concept of dual mobility socket has been used, for increase stability and mobility. The first mobility is between the femoral head and the polyethylene insert, and the second mobility between the polyethylene and the metal cup socket. The cup, covered with alumina was impacted without cement. The purpose of this study was to evaluate the long-term results of a retrospective series of primary arthroplasty with this cup and a cemented Charnley type femoral component.

Materiel and methods: This study included 437 hip replacements performed between 1984 and 1990, in 389 patients with osteoarthritis (62.5%), dysplasia (11.4%), necrosis (8%), rheumatoid arthritis (5%), post-traumatic and others (13.1%). The average age of the patients at the time of the index procedure was 61 years. The patients were followed at routine intervals for clinical and radiographic review. The clinical results were appreciated with Postel–Merle d’Aubigne score (PMA). Osteolysis was noted according to Gruen zones.

Results: The outcome is known for 345 hips (79%): 164 alive without revision at a mean of 16,5 years follow-up, 137 died without revision, and 44 failures. 92 (21%) were lost at follow up. Revision, for aseptic loosening of femoral or acetabular component, was performed in 30 hips(6.8%). 5 dislocation occurred and were revised: 2 early regarding to technical errors and 3 after 10 years or more of follow up. 7 hips were revised for deep infection, and 2 for unaccountable pain.

At more than 15 years follow-up, the mean PMA hip functional score was 17.1. 66% of the hips was A, 4% were B and 30% were C according to the Charnley’s score. Of the 164 hips in the patients who had survived at least 15 years, 28 had femoral osteolysis in zone 7, and 31 had femoral osteolysis in one (or more) of the other 13 zones. 6 hips had an impingement sign on the neck of the femoral component, without aseptic loosening.

According to Kaplan-Meier analysis, the fifteen year survival rate, was 84.36% with revision for any reason (infection, dislocation, osteolysis…) for end point. The young age of the patients at the time of the index surgery is correlated with loosening.

Conclusion: The long term results of the press fit, double mobility socket with cemented Charnley type femoral stem are good. The prevalence of revision for dislocation is very low in our series. But this concept do not avoid wear, osteolysis and aseptic loosening, specially in young active patients. We can recommand this type of prosthesis for patients over 70 years, and/or with high risk of dislocation.


Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 256 - 256
1 Jul 2008
LECLERCQ S
Full Access

Purpose of the study: The stability of the patella over the femur depends on several factors, one being the rotatory freedom of the tibia. Femorotibial rotatory laxity of a total knee arthroplasty (TKA) can be:

dictated by the congruence of the polyethylene in an ultracongruent plateau;

completely free, depending solely on the ligament structures in a self-aligning prosthesis;

the consequence of a compromise between the two, using a semi-constrained prosthesis. With the OMNIA system, the unique femoral piece can be combined with a Wallaby ultracongruent plateau (WUC), a self-aligning mobile plateau (SAL), or a semiconstrained plateau with preservation of the posterior cruciate ligament (Wallaby 1, W1). A comparative study of these three prostheses was performed to evaluate the influence of femorotibial rotatory constraint on patellar stability.

Material and methods: The series included 157 TKA: 68 SAL,44 WUC and 45 W1. The same surgeon operated all patients in three successive series. The operative technique was the same using the same instrument set. The femoral piece was inserted first with an automatic rotation systematically set at 5° with the posterior condyles. The knees were varus (71%), valgus (20%) and perfectly aligned (9%). A patellar prosthesis was used in only 10.8% due to excessive wear as assessed intraoperatively. A medial approach was used for the varus knees and a lateral approach for the valgus knees greater than 10°, independently of preoperative patellar stability. Radiological outcome was assessed on the 30° femoro-patellar views. Patellae tilted more than 3° and/or offset more than 5 mm were considered to be excentered.

Results: In the SAL group, 55 patellae were centered and 13 off-centered. The ratio was 36 for 6 in the WUC group and 39 for 6 in the W1 group. The percentages of centered patellae were respectively 80, 85, and 86%. The difference was not significant (p=0.66. For knees with preoperative genu varum, the percentages of centered patella were 83, 84, and 86%; and for genu varum, 72, 0 and 75% (sample too small for statistical significance).

Discussion: The use of rotatory constraint for TKA does not affect patellar stability. This was confirmed in a subpopulation of genu varum knees and not in genu valgum.