Total hip arthroplasty is well established as a successful treatment modality for end stage arthritis, with a variety of components currently available. However, utilising traditional stemmed implants in patients with distorted proximal femoral geometry can be technically challenging with increased risk of complications. We present seven patients with distorted proximal femoral anatomy or failed hip arthroplasty in whom a technically challenging primary or revision operation was simplified by use of a Proxima stem. This is a short, stemless, metaphyseal loading implant with a pronounced lateral flare. At twelve months follow up there have been no complications with average improvement in Oxford and Harris scores of forty and forty-nine respectively. Radiological analysis shows all stems to be stable and well fixed. Designed primarily as a bone conserving implant for primary hip arthroplasty we propose that the Proxima prosthesis also be considered in cases where a conventional stemmed implant may not be suitable due to challenging proximal femoral anatomy. The use of the stemless Proxima implant provided a simple solution in seven difficult and potentially lengthy complex primary and revision hip arthroplasties with gratifying clinical and radiological outcomes.
Previous research has shown an increase in chromosomal aberrations in patients with worn implants. The type of aberration depended on the type of metal alloy in the prosthesis. We have investigated the metal-specific difference in the level of DNA damage (DNA stand breaks and alkali labile sites) induced by culturing human fibroblasts in synovial fluid retrieved at revision arthroplasty. All six samples from revision cobalt-chromium metal-on-metal and four of six samples from cobalt-chromium metal-on-polyethylene prostheses caused DNA damage. By contrast, none of six samples from revision stainless-steel metal-on-polyethylene prostheses caused significant damage. Samples of cobalt-chromium alloy left to corrode in phosphate-buffered saline also caused DNA damage and this depended on a synergistic effect between the cobalt and chromium ions. Our results further emphasise that epidemiological studies of orthopaedic implants should take account of the type of metal alloy used.
This study investigates the use of porous biphasic ceramics as graft extenders in impaction grafting of the femur during revision hip surgery. Impaction grafting of the femur was performed in four groups of sheep. Group one received pure allograft, group two 50% allograft and 50% BoneSave, group three 50% allograft and 50% BoneSave type 2 and group four 10% allograft and 90% BoneSave as the graft material. Function was assessed using an index of pre- and post-operative peak vertical ground reaction force ratios. Changes in bone mineral density were measured by dual energy X ray absorptiometry (DEXA) scanning. Loosening and subsidence were assessed radiographically and by histological examination of the explanted specimens. There was no statistically significant difference between the four groups after 18 months of unrestricted functional loading for all outcome measures.
The long-term biological effects of wear debris are unknown. We have investigated whether there is any evidence of cumulative mutagenic damage in peripheral blood lymphocytes of patients undergoing revision arthroplasty of predominantly metal-on-plastic total hip replacements compared with those at primary arthroplasty. There was a threefold increase in aneuploidy and a twofold increase in chromosomal translocations which could not be explained by the confounding variables of smoking, gender, age and diagnostic radiographs. In the patients with TiVaAl prostheses there was a fivefold increase in aneuploidy but no increase in chromosomal translocations. By contrast, in patients with cobalt-chrome prostheses there was a 2.5-fold increase in aneuploidy and a 3.5-fold increase in chromosomal translocations. In six patients with stainless-steel prostheses there was no increase in either aneuploidy or chromosomal translocations. Our results suggest that future epidemiological studies of the putative long-term risks of joint replacement should take into account the type of alloy used in the prosthesis.