Femoral stem design affects periprosthetic bone mineral density (BMD), which may impact long term survival of cementless implants in total hip arthroplasty (THA). The aim of this study was to examine proximal femoral BMD in three morphologically different uncemented femoral stems designs to investigate whether one particular design resulted in improved preservation of BMD. 119 patients were randomized to receive either a proximally coated dual taper wedge stem, a proximally coated anatomic stem or a fully coated collarless triple tapered stem. Dual energy x-ray absorptiometry scans (Lunar iDXA, GE Healthcare, Madison, WI) assessed BMD across the seven Gruen zones pre-operatively, and post-operatively at 6-weeks, 1-year, and 2-years comparing the unoperated contralateral femur as a control. BMD increased in zones one (2.5%), two (17.1%), three (13.0%), five (10%) and six (17.9%) for all stems. Greater preservation of BMD was measured on the lateral cortex (zone 2) for both the dual taper wedge and anatomic stems (p = 0.019). The dual taper wedge stem also demonstrated preservation of BMD in the medial calcar (zone 7) whilst the anatomic and triple taper stem declined in this region, however this was not statistically significant (p = 0.059). BMD decreased on average by 2.1% in the mid-diaphysis region, distal to the stem tip (zone 4) for all implants. All stems performed equivalently at final follow-up in all patient reported outcome measures. Implant design was shown to impact BMD changes. All stems preserved bone at the metaphyseal level in early follow-up, which theoretically indicates favorable implant fixation. There was no significant stress shielding observed, however longer follow-up is required to elucidate the impact of this finding on implant survivorship.