Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_11 | Pages 83 - 83
1 Dec 2020
Shah DS Taylan O Labey L Scheys L
Full Access

Understanding the long-term effects of total knee arthroplasty (TKA) on joint kinematics is vital to assess the success of the implant design and surgical procedure. However, while in vitro cadaveric studies quantifying post-operative biomechanics primarily reflect joint behaviour immediately after surgery,1in vivo studies comprising of follow-up TKA patients often reflect joint behaviour a few months after surgery.2 Therefore, the aim of this cadaveric study was to explore the long-term effects of TKA on tibiofemoral kinematics of a donor specimen, who had already undergone bilateral TKA, and compare them to post-operative kinematics reported in the literature.

Two fresh-frozen lower limbs from a single donor (male, age: 83yr, ht: 1.83m, wt: 86kg), who had undergone bilateral TKA (Genesis II, Smith&Nephew, Memphis, USA) 19 years prior to his demise, were obtained following ethical approval from the KU Leuven institutional board. The specimens were imaged using computed tomography (CT) and tested in a validated knee simulator3 replicating active squatting and varus-valgus laxity tests. Tibiofemoral kinematics were recorded using an optical motion capture system and compared to various studies in the literature using the same implant – experimental studies based on cadaveric specimens (CAD)1,4 and an artificial specimen (ART)5, and a computational study (COM)6.

Maximum tibial abduction during laxity tests for the left leg (3.54°) was comparable to CAD (3.30°), while the right leg exhibited much larger joint laxity (8.52°). Both specimens exhibited valgus throughout squatting (left=2.03±0.57°, right=5.81±0.19°), with the change in tibial abduction over the range of flexion (left=1.89°, right=0.64°) comparable to literature (CAD=1.28°, COM=2.43°). The left leg was externally rotated (8.00±0.69°), while the right leg internally rotated (−15.35±1.50°), throughout squatting, with the change in tibial rotation over the range of flexion (left=2.61°, right=4.79°) comparable to literature (CAD=5.52°, COM=4.15°). Change in the femoral anteroposterior translation over the range of flexion during squatting for both specimens (left=14.88mm, right=6.76mm) was also comparable to literature (ART=13.40mm, COM=20.20mm).

Although TKA was reportedly performed at the same time on both legs of the donor by the same surgeon, there was a stark difference in their post-operative joint kinematics. A larger extent of intraoperative collateral ligament release could be one of the potential reasons for higher post-operative joint laxity in the right leg. Relative changes in post-operative tibiofemoral kinematics over the range of squatting were similar to those reported in the literature. However, differences between absolute magnitudes of joint kinematics obtained in this study and findings from the literature could be attributed to different surgeons performing TKA, with presumable variations in alignment techniques and/or patient specific instrumentation, and the slightly dissimilar ranges of knee flexion during squatting.

In conclusion, long-term kinematic effects of TKA quantified using in vitro testing were largely similar to the immediate post-operative kinematics reported in the literature; however, variation in the behaviour of two legs from the same donor suggested that intraoperative surgical alterations might have a greater effect on joint kinematics over time.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_11 | Pages 31 - 31
1 Dec 2020
Shah DS Taylan O Berger P Labey L Vandenneucker H Scheys L
Full Access

Orthopaedic training sessions, vital for surgeons to understand post-operative joint function, are primarily based on passive and subjective joint assessment. However, cadaveric knee simulators, commonly used in orthopaedic research,1 could potentially benefit surgical training by providing quantitative joint assessment for active functional motions. The integration of cadaveric simulators in orthopaedic training was explored with recipients of the European Knee Society Arthroplasty Travelling Fellowship visiting our institution in 2018 and 2019. The aim of the study was to introduce the fellows to the knee joint simulator to quantify the surgeon-specific impact of total knee arthroplasty (TKA) on the dynamic joint behaviour, thereby identifying potential correlations between surgical competence and post-operative biomechanical parameters.

Eight fellows were assigned a fresh-frozen lower limb each to plan and perform posterior-stabilised TKA using MRI-based patient-specific instrumentation. Surgical competence was adjudged using the Objective Structured Assessment of Technical Skills (OSATS) adapted for TKA.2 All fellows participated in the in vitro specimen testing on a validated knee simulator,3 which included motor tasks – passive flexion (0°-120°) and active squatting (35°-100°) – and varus-valgus laxity tests, in both the native and post-operative conditions. Tibiofemoral kinematics were recorded with an optical motion capture system and compared between native and post-operative conditions using a linear mixed model (p<0.05). The Pearson correlation test was used to assess the relationship between the OSATS scores for each surgeon and post-operative joint kinematics of the corresponding specimen (p<0.05).

OSATS scores ranged from 79.6% to 100% (mean=93.1, SD=7.7). A negative correlation was observed between surgical competence and change in post-operative tibial kinematics over the entire range of motion during passive flexion – OSATS score vs. change in tibial abduction (r=−0.87; p=0.003), OSATS score vs. change in tibial rotation (r=−0.76; p=0.02). When compared to the native condition, post-operative tibial internal rotation was higher during passive flexion (p<0.05), but lower during squatting (p<0.033). Post-operative joint stiffness was greater in extension than in flexion, without any correlation with surgical competence.

Although trained at different institutions, all fellows followed certain standard intraoperative guidelines during TKA, such as achieving neutral tibial abduction and avoiding internal tibial rotation,4 albeit at a static knee flexion angle. However, post-operative joint kinematics for dynamic motions revealed a strong correlation with surgical competence, i.e. kinematic variability over the range of passive flexion post-TKA was lower for more skilful surgeons. Moreover, actively loaded motions exhibited stark differences in post-operative kinematics as compared to those observed in passive motions. In vitro testing on the knee simulator also introduced the fellows to new quantitative parameters for post-operative joint assessment.

In conclusion, the inclusion of cadaveric simulators replicating functional joint motions could help quantify training paradigms, thereby enhancing traditional orthopaedic training, as was also the unanimous opinion of all participating fellows in their positive feedback.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1466 - 1470
1 Oct 2010
Didden K Luyckx T Bellemans J Labey L Innocenti B Vandenneucker H

The biomechanics of the patellofemoral joint can become disturbed during total knee replacement by alterations induced by the position and shape of the different prosthetic components. The role of the patella and femoral trochlea has been well studied. We have examined the effect of anterior or posterior positioning of the tibial component on the mechanisms of patellofemoral contact in total knee replacement. The hypothesis was that placing the tibial component more posteriorly would reduce patellofemoral contact stress while providing a more efficient lever arm during extension of the knee.

We studied five different positions of the tibial component using a six degrees of freedom dynamic knee simulator system based on the Oxford rig, while simulating an active knee squat under physiological loading conditions. The patellofemoral contact force decreased at a mean of 2.2% for every millimetre of posterior translation of the tibial component. Anterior positions of the tibial component were associated with elevation of the patellofemoral joint pressure, which was particularly marked in flexion > 90°.

From our results we believe that more posterior positioning of the tibial component in total knee replacement would be beneficial to the patellofemoral joint.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 737 - 742
1 May 2010
Verlinden C Uvin P Labey L Luyckx JP Bellemans J Vandenneucker H

Malrotation of the femoral component is a cause of patellofemoral maltracking after total knee arthroplasty. Its precise effect on the patellofemoral mechanics has not been well quantified. We have developed an in vitro method to measure the influence of patellar maltracking on contact. Maltracking was induced by progressively rotating the femoral component either internally or externally. The contact mechanics were analysed using Tekscan. The results showed that excessive malrotation of the femoral component, both internally and externally, had a significant influence on the mechanics of contact. The contact area decreased with progressive maltracking, with a concomitant increase in contact pressure. The amount of contact area that carries more than the yield stress of ultra-high molecular weight polyethylene significantly increases with progressive maltracking. It is likely that the elevated pressures noted in malrotation could cause accelerated and excessive wear of the patellar button.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 5 | Pages 683 - 690
1 May 2009
Victor J Van Doninck D Labey L Van Glabbeek F Parizel P Bellemans J

The understanding of rotational alignment of the distal femur is essential in total knee replacement to ensure that there is correct placement of the femoral component. Many reference axes have been described, but there is still disagreement about their value and mutual angular relationship. Our aim was to validate a geometrically-defined reference axis against which the surface-derived axes could be compared in the axial plane. A total of 12 cadaver specimens underwent CT after rigid fixation of optical tracking devices to the femur and the tibia. Three-dimensional reconstructions were made to determine the anatomical surface points and geometrical references. The spatial relationships between the femur and tibia in full extension and in 90° of flexion were examined by an optical infrared tracking system.

After co-ordinate transformation of the described anatomical points and geometrical references, the projection of the relevant axes in the axial plane of the femur were mathematically achieved. Inter- and intra-observer variability in the three-dimensional CT reconstructions revealed angular errors ranging from 0.16° to 1.15° for all axes except for the trochlear axis which had an interobserver error of 2°. With the knees in full extension, the femoral transverse axis, connecting the centres of the best matching spheres of the femoral condyles, almost coincided with the tibial transverse axis (mean difference −0.8°, sd 2.05). At 90° of flexion, this femoral transverse axis was orthogonal to the tibial mechanical axis (mean difference −0.77°, sd 4.08). Of all the surface-derived axes, the surgical transepicondylar axis had the closest relationship to the femoral transverse axis after projection on to the axial plane of the femur (mean difference 0.21°, sd 1.77). The posterior condylar line was the most consistent axis (range −2.96° to −0.28°, sd 0.77) and the trochlear anteroposterior axis the least consistent axis (range −10.62° to +11.67°, sd 6.12). The orientation of both the posterior condylar line and the trochlear anteroposterior axis (p = 0.001) showed a trend towards internal rotation with valgus coronal alignment.