Vol. 99-B, Issue SUPP_20 | Pages 22 - 22
1 Dec 2017
Wong AKC Kumta BSM Sze CL
Full Access

Navigation-assisted surgery has been reported to enhance resection accuracy in bone sarcoma surgery. Patient-specific instruments (PSIs) have been proposed as a simpler alternative with fewer setup facilities. We investigated the use of 3D surgical planning and PSI in realising computer planning of complex resections in bone sarcoma patients with regards to surgical accuracy, problems, and early clinical results.

We retrospectively studied twelve patients with bone sarcoma treated surgically by PSIs with 3D planning. The procedure was planned using engineering software. The resection accuracy was accessed by comparing CT images of tumour specimens with the planned in seven patients. Mean age was 30.9 (9 – 64). Mean follow-up was 3.1 year (0.5 – 5.3).

31 planes of bone resections were successfully performed using the technique and were considered accurate. The mean time required for placing PSIs was 5.7 minutes (1 – 10) and performing bone osteotomies with the assistance of PSIs was 4.7 minutes (2 – 7). The mean maximum deviation error was 1.7mm (0.5 – 4.4). One PSI was broken during bone resection, and one patient needed re-resection using the same PSI. One pelvic patient died of local recurrence and lung metastases six months postoperatively. One patient developed a soft tissue local recurrence and lung metastasis at 20 months after surgery. The mean MSTS functional score was 27.9 (21 – 30). There were no complications related to 3D planning and PSIs.

In selected patients, 3D surgical planning and PSIs replicate complex bone resections and reconstructions in bone sarcoma surgery. Comparative studies with conventional or navigation- assisted resections are required.