The prosthetic impingement occurs if the range of motion of the hip exceeds implant mobility or in case of component malorientation. This retrieval study was designed to assess the frequency and the risk factors of this phenomenon. The frequency and the severity of the impingement were calculated from a continuous series of 311 cups retrievals collected between 1989 and 2004 by a single surgeon. The reason for retrieval was loosening (131 cases), infection (43 cases), instability (56 cases), osteolysis (28 cases), unexplained pain (48 cases) and prosthetic impingent (5 cases all with hard bearings). The notching at the cup rim was assessed twice by two examiners with optic magnification. The risk factors were analyzed from clinical charts by univariate and cox multihazard.Introduction
Material and Methods