At 70 patients the femur was augmented (mesh, cerclage(s) and/or plate). A short stem (115–135mm) was used in 48% and a long stem (165–225mm) in 52% of the cases.
Postoperative the femoral stem had a varus position (6 –10°) in 4 cases. In 6 cases the stem was in posterior and 3 cases in anterior position. Progressive subsidence was demonstrated in 1 patient increasing to 5 mm at 1 year fu. Three patients showed subsidence of 3 mm at 3 months, but were stable afterwards. Slightly increasing radiolucent lines (>
2 mm) were noticed in 2 patients in subcortical areas, but never any signs of osteolysis. Intra- and postoperative femoral fractures occurred in 12 patients (13,6%). There were 3 dislocations and 1 loosening of a trochanteric osteotomy. Despite the fractures there was no loosening of a stem during follow-up period. After bivariate multilevel analysis, conducted to determine a relationship between the fracture complication and all other variables, a multivariate model was developed of the most significant variables to determine the predictor factors for femoral fracture. Only the Endo-classification and age are predictors for a fracture following a RIG procedure (p-value 0.003). Each higher EC type increases the probability of a fracture after RIG 2.01 times and each added year of life 1.07 times. So the risk of getting a femoral fracture increases exponentially with age from 1.7% at 55 years to 15.5% at 90 years in EC type 1. The fracture risk in EC type 4 increases from 4.7% at 40 years to almost 60% at 90 years.
a proper randomized controlled trial; based on a total hip prosthesis with or without hydroxyapatite-coating with one identical geometry; primary uncemented total hip arthroplasty; and used objective, validated clinical and radiographic outcome measurements.