14 Nov 2024
Mischler D Kessler F Zysset P Varga P
Full Access

Introduction

Pedicle screw loosening in posterior instrumentation of thoracolumbar spine occurs up to 60% in osteoporotic patients. These complications may be alleviated using more flexible implant materials and novel designs that could be optimized with reliable computational modeling. This study aimed to develop and validate non-linear homogenized finite element (hFE) simulations to predict pedicle screw toggling.

Method

Ten cadaveric vertebral bodies (L1-L5) from two female and three male elderly donors were scanned with high-resolution peripheral quantitative computed tomography (HR-pQCT, Scanco Medical) and instrumented with pedicle screws made of carbon fiber-reinforced polyether-etherketone (CF/PEEK). Sample-specific 3D-printed guides ensured standardized instrumentation, embedding, and loading procedures. The samples were biomechanically tested to failure in a toggling setup using an electrodynamic testing machine (Acumen, MTS) applying a quasi-static cyclic testing protocol of three ramps with exponentially increasing peak (1, 2 and 4 mm) and constant valley displacements. Implant-bone kinematics were assessed with a stereographic 3D motion tracking camera system (Aramis SRX, GOM). hFE models with non-linear, homogenized bone material properties including a strain-based damage criterion were developed based on intact HR-pQCT and instrumented 3D C-arm scans. The experimental loading conditions were imposed, the maximum load per cycle was calculated and compared to the experimental results. HR-pQCT-based bone volume fraction (BV/TV) around the screws was correlated with the experimental peak forces at each displacement level.