Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 51 - 51
1 Feb 2017
Bragdon C Barr C Berry D Della Valle C Garvin K Johanson P Clohisy J Malchau H
Full Access

Introduction

The first highly crosslinked and melted polyethylene acetabular component for use in total hip arthroplasty was implanted in 1998 and femoral heads larger than 32mm in diameter introduced 2004.

The purpose of this study was to re-assemble a previous multi-center patient cohort in order to evaluate the radiographic and wear analysis of patients receiving this form of highly crosslinked polyethylene articulating against large diameter femoral heads at a minimum of 10 years follow-up.

Methods

Two centers contributed patients to this ongoing clinical study. Inclusion criteria for patients was: primary THR; femoral heads greater than 32mm; minimum 10 year follow-up. 69 hips have been enrolled with an average follow-up of 11.2 years (10–15), 32 females (50%). Wear analysis was performed using the Martell Hip Analysis software. Radiographic grading was performed on the longest follow-up AP hip films. The extent of radiolucency in each zone greater than 0.5mm in thickness was recorded along with the presence of sclerotic lines and osteolysis.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 52 - 52
1 Feb 2017
Bragdon C Barr C Berry D Della Valle C Garvin K Johanson P Clohisy J Malchau H
Full Access

Introduction

The first highly crosslinked and melted polyethylene acetabular component for use in total hip arthroplasty was implanted in 1998. Numerous publications have reported reduced wear rates and a reduction in particle induced peri-prosthetic osteolysis at short to mid-term follow-up.

The purpose of this study was to re-assemble a previous multi-center patient cohort in order to evaluate the radiographic and wear analysis of patients receiving this form of highly crosslinked polyethylene articulating against 32mm femoral heads or less at a minimum of 13 years follow-up.

Methods

Inclusion criteria for patients was a primary THR with femoral heads 32mm or less and a minimum 13 year follow-up. 139 hips have been enrolled with an average follow-up of 13.7 years (13–16), 80 females (57%). Wear analysis was performed using the Martell Hip Analysis software. Radiographic grading was performed on the longest follow-up AP hip films. The extent of radiolucency in each zone greater than 0.5mm in thickness was recorded along with the presence of sclerotic lines and osteolysis.