The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 1012 - 1019
1 Aug 2017
Howard DP Wall PDH Fernandez MA Parsons H Howard PW

Aims

Ceramic-on-ceramic (CoC) bearings in total hip arthroplasty (THA) are commonly used, but concerns exist regarding ceramic fracture. This study aims to report the risk of revision for fracture of modern CoC bearings and identify factors that might influence this risk, using data from the National Joint Registry (NJR) for England, Wales, Northern Ireland and the Isle of Man.

Patients and Methods

We analysed data on 223 362 bearings from 111 681 primary CoC THAs and 182 linked revisions for bearing fracture recorded in the NJR. We used implant codes to identify ceramic bearing composition and generated Kaplan-Meier estimates for implant survivorship. Logistic regression analyses were performed for implant size and patient specific variables to determine any associated risks for revision.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 585 - 588
1 May 2016
Bayley E Brown S Bhamber NS Howard PW

Aims

The place of thromboprophylaxis in arthroplasty surgery remains controversial, with a challenging requirement to balance prevention of potentially fatal venous thrombo-embolism with minimising wound-related complications leading to deep infection. We compared the incidence of fatal pulmonary embolism in patients undergoing elective primary total hip arthroplasty (THA) between those receiving aspirin, warfarin and low molecular weight heparin (LMWH) for the chemical component of a multi-modal thromboprophylaxis regime.

Patients and Methods

A prospective audit database was used to identify patients who had died within 42 and 90 days of surgery respectively between April 2000 and December 2012. A case note review was performed to ascertain the causes of death.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1000 - 1004
1 Aug 2008
Wraighte PJ Howard PW

Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless, polished, tapered femoral stem between 1992 and 1998.

The mean follow-up period was 10.5 years (6.3 to 14.1). The median pre-operative bone defect score was 3 (interquartile range (IQR) 2 to 3) using the Endo-Klinik classification.

The median subsidence at one year post-operatively was 2 mm (IQR 1 to 3). At the final review the median Harris hip score was 80.6 (IQR 67.6 to 88.9) and the median subsidence 2 mm (IQR 1 to 4). Incorporation of the allograft into trabecular bone and secondary remodelling were noted radiologically at the final follow-up in 87% (393 of 452 zones) and 40% (181 of 452 zones), respectively.

Subsidence of the Exeter stem correlated with the pre-operative Endo-Klinik bone loss score (p = 0.037). The degree of subsidence at one year had a strong association with long-term subsidence (p < 0.001). There was a significant correlation between previous revision surgery and a poor Harris Hip score (p = 0.028), and those who had undergone previous revision surgery for infection had a higher risk of complications (p = 0.048). Survivorship at 10.5 years with any further femoral operation as the end-point was 92% (95% confidence interval 82 to 97).


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 12 | Pages 1611 - 1616
1 Dec 2005
Badhe NP Howard PW

We evaluated the use of a stemmed acetabular component in the treatment of severe acetabular deficiency during revision and complex primary total hip arthroplasty.

There were 31 hips of which 24 were revisions (20 for aseptic loosening, four for infection) and the remainder were complex primary arthroplasties. At a mean follow-up of 10.7 years (6 to 12.8), no component had been revised for aseptic loosening; one patient had undergone a revision of the polyethylene liner for wear. There was one failure because of infection. At the latest follow-up, the cumulative survival rate for aseptic loosening, with revision being the end-point, was 100%; for radiographic loosening it was 92% and for infection and radiographic loosening it was 88%. These results justify the continued use of this stemmed component for the reconstruction of severe acetabular deficiency.