Golf is considered low-impact sport, but concerns exist about whether golf swing can be performed in safe manner after THA. The purpose of this study was to clarify dynamic hip kinematics during golf swing after THA using image-matching techniques. This study group consisted of eight right-handed recreational golfers with 10 primary THAs. Each operation was performed using a posterolateral approach with combined anteversion technique. Nine of ten polyethylene liners used had elevated portion of 15°. Continuous radiographic images of five trail and five lead hips during golf swing were recorded using a flat panel X-ray detector (Fig. 1) and analyzed using image-matching techniques (Fig. 2). The relative distance between the center of cup and femoral head and the minimum liner-to-stem distance were measured using a CAD software program. The cup inclination, cup anteversion, and stem anteversion were measured in postoperative CT data. Hip kinematics, orientation of components, and cup-head distance were compared between patients with and without liner-to-stem contact by Mann-Whitney INTRODUCTION
METHODS
Controversy still exists as to whether total knee arthroplasty (TKA) provides reproducible knee kinematics during activities. In this study, we evaluated the A total of twenty four knees in nineteen patients following cruciate-retaining (CR) or posterior-stabilized (PS) TKA were randomly included in the study. The twenty-four knees included 22 female knees and 2 male knees in patients aged 73 years. The pre-operative diagnosis was osteoarthritis in 22 knees and rheumatoid arthritis in 2 knees. The average follow-up period after surgery was 29 months, and average post-operative knee extension/flexion angle was 2°/121°. The average knee score was 93 and the average functional score was 77. Continuous sagittal radiological images were obtained during stair-climbing for each patient using a large flat panel detector. Anteroposterior (AP) tibiofemoral position, implant flexion, and axial rotation angles were determined in three dimensions using a 3D-to-2D model-to-image registration technique. In CR TKA, the minimum distances between the femoral trochlea and the intercondylar eminence of the tibial insert were measured using a CAD software program. In PS TKA, the minimum distances between the femoral cam and the posterior aspect of the tibial post and between the femoral trochlea and the anterior aspect of the tibial post were measured.Introduction
Patients and Methods
3D-to-2D model registration technique has been used for evaluating 3D kinematics from 3D surface models of the prostheses or bones and radiographic image sequences. However, no studies have employed these techniques to evaluate Dynamic hip kinematics during gait, squatting, chair-rising, and twisting were analyzed for six healthy subjects and eleven patients with osteoarthritis (OA). Continuous anteroposterior radiographic images were recorded using a flat panel X-ray detector Introduction
Measurement
This study presents the use of precision surface machining on artificial joint bearing surfaces in order to inhibit macrophage activation. Ultra-high molecular weight polyethylene (UHMWPE) is widely used as a bearing material in polymer-on-hard joint prostheses. However, UHMWPE wear particles are considered to be a major factor in long-term osteolysis and implant loosening. Several studies report that wear particle size is a critical factor in macrophage activation, with particles in the size range of 0.1 – 1.0 μm being the most biological active. The surface for a conventional Co-Cr-Mo alloy joint implant generally has a 10.0 – 20.0 nm roughness. After precision machining, the Co-Cr-Mo alloy surface had a 1.0 – 2.0 nm roughness with scattered concave shapes up to 50 nm in depth. This precision surface machining method used a typical lapping method, but the relationship between the slurry and the machining surface was strictly controlled in order to emphasize the micro-erosion mechanism. A pin-on-disc wear tester capable of multidirectional motion was used to verify that the new surface was the most appropriate for joints. Tests were carried out in 25% (v/v) fetal calf serum with sodium azide to retard bacterial growth. UHMWPE pins, 12.0 mm in diameter with a mean molecular weight of 6.0 million, were placed on the Co-Cr-Mo alloy disc at a contact pressure of 6.0 MPa. A sliding speed of 12.1 mm/s, and a total sliding distance of 15.0 km were applied. The new surface reduced the amount of UHMWPE wear, which would ensure the long-term durability of joints. The new surface also enlarged the size of UHMWPE particles, but did not change their morphological aspect. Primary human peripheral blood mononuclear phagocytes were cultured with the particles. The wear particles generated on the new surface inhibited the production of IL-6, which indicates a reduction of induced tissue reaction and joint loosening.