Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 34 - 34
1 Dec 2021
Elkington R Beadling A Hall R Pandit H Bryant M
Full Access

Abstract

Objectives

Current use of hard biomaterials such as cobalt-chrome alloys or ceramics to articulate against the relatively soft, compliant native cartilage surface reduces the joint contact area by up to two thirds. This gives rise to high and abnormal loading conditions which promotes degradation and erosion of the mating cartilage leading to pain, stiffness, and loss of function. Biomimetic soft lubrication strategies have been developed by grafting hydrophilic polymers onto substrates to form a gel-type surface. Surface grafted gels mimic the natural mechanisms of friction dissipation in synovial joints, showing a promising potential for use in hemiarthroplasty. This project aims to develop implant surfaces with properties tailored to match articular cartilage to retain and promote natural joint function ahead of total joint replacement.

Methods

Four different types of monomers were grafted in a one-step photopolymerisation procedure onto polished PEEK substrates. The functionalised surfaces were investigated using surface wettability, FTIR, and simplified 2D-tribometry tests against glass and animal cartilage specimens to assess their lubricity and mechanical properties for hemiarthroplasty articulations.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_2 | Pages 42 - 42
1 Jan 2019
Lal S Hall R Tipper JL
Full Access

Since 2010, there has been a sharp decline in the use of metal-on-metal joint replacement devices due to adverse responses associated with the release of metal wear particles and ions in patients. Surface engineered coatings offer an innovative solution to this problem by covering metal implant surfaces with biocompatible and wear resistant materials. The present study tests the hypothesis whether surface engineered coatings can reduce the overall biological impact of a device by investigating recently introduced silicon nitride coatings for joint replacements. Biological responses of peripheral blood mononuclear cells (PBMNCs) to Si3N4 model particles, SiNx coating wear particles and CoCr wear particles were evaluated by testing cytotoxicity, inflammatory cytokine release, oxidative stress and genotoxicity.

Clinically relevant wear particles were generated from SiNx-on-SiNx and CoCr-on-CoCr bearing combinations using a multidirectional pin-on-plate tribometer. All particles were heat treated at 180°C for 4 h to destroy endotoxin contamination. Whole peripheral blood was collected from healthy donors (ethics approval BIOSCI 10–108, University of Leeds). The PBMNCs were isolated using Lymphoprep (Stemcell) and incubated with particles at various volumetric concentrations (0.5 to 100 µm3 particles/cell) for 24 h in 5% (v/v) CO2 at 37°C. After incubation, cell viability was measured using the ATPlite assay (Perkin Elmer); TNF-alpha release was measured by ELISA (Invitrogen); oxidative stress was measured using H2DCFDA (Abcam); and DNA damage was measured by comet assay (Tevigen). The results were expressed as mean ± 95% confidence limits and the data was analysed using one-way ANOVA and Tukey-Kramer post-hoc analysis.

No evidence of cytotoxicity, oxidative stress, TNF-alpha release, or DNA damage was observed for the silicon nitride particles at any of the doses. However, CoCr wear particles caused cytotoxicity, oxidative stress, TNF-alpha release and DNA damage in PBMNCs at high doses (50 µm3 particles per cell). This study has demonstrated the in-vitro biocompatibility of SiNx coatings with primary human monocytic cells. Therefore, surface engineered coatings have potential to significantly reduce the biological impact of metal components in future orthopaedic devices.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_16 | Pages 42 - 42
1 Oct 2016
Pasko K Hall R Neville A Tipper J
Full Access

Surgical interventions for the treatment of chronic neck pain, which affects 330 million people globally [1], include fusion and cervical total disc replacement (CTDR). Most of the currently clinically available CTDRs designs include a metal-on-polymer (MoP) bearing. Numerous studies suggest that MoP CTDRs are associated with issues similar to those affecting other MoP joint replacement devices, including excessive wear and wear particle-related inflammation and osteolysis [2,3]. A device with a metal-on-metal (MoM) bearing has been investigated in the current study.

Six MoM CTDRs made from high carbon cobalt-chromium (CoCr) were tested in a six-axis spine simulator, under standard ISO testing protocol (ISO-18192-1) for a duration of 4 million cycles (MC). Foetal bovine calf serum (25%v/v), used as a lubricant, was changed every 3.3×105 cycles and saved for particle analysis. Components were taken down for measurements after each 106 cycles; surface roughness, damage modes and gravimetric wear were assessed.

The mean wear rate of the MoM CTDRs was 0.24mm3/MC (SD=0.03), with the total volume of 0.98mm3 (SD=0.01) lost over the test duration. Throughout the test, the volumetric wear was linear; no significant bedding-in period was observed. The mean pre-test surface roughness decreased from 0.019μm (SD=0.005) to 0.012μm (SD=0.002) after 4MC of testing. Prior to testing, fine polishing marks on the bearing surfaces were observed using light microscopy. Following 4MC of testing, these polishing marks had been removed. Consistently across all components, surface discolouration and multidirectional, criss-crossing, circular wear tracks, caused by abrasive wear, were observed.

The wear results showed low wear rates exhibited by MoM CTDRs (0.24mm3/MC), when compared CTDR designs incorporating metal-on-polymer bearings (0.56mm3/MC) [4] as well as MoM lumbar CTDRs [5,6] (0.76mm3/MC – 6.2mm3/MC). These findings suggest that MoM CTDRs are more wear resistant than MoP CTDRs, however the particle characterisation and biological consequences of wear remain to be determined.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 74 - 74
1 Jul 2014
Brandolini N Kapur N Hall R
Full Access

Summary Statement

Burst fractures were simulated in vitro on human cadaveric spine segments. Displacement of the facet joints and pedicles were measured throughout the fracture process showing how these bony structures behave when an impact load is delivered.

Introduction

Burst fractures account for almost 30% of all spinal injuries, which may result in severe neurological deficit, spinal instability and hence life impairment1. The onset of the fracture is usually traumatic, caused by a high-energy impact loading. Comminution of the endplates and vertebral body, retropulsion of fragments within the canal and increase of the intrapedicular distance are typical indicators of the injury. Experimental and numerical studies have reported strain concentration at the base of the pedicles, suggesting that the posterior processes play a fundamental role in the fracture initiation2,3. However, little is known about the dynamic behaviour of the vertebra undergoing an impact load. The aim of this study was to provide an in vitro cadaveric investigation on burst fracture, focusing on the widening of the facet joints and pedicles during the fracture development.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 182 - 182
1 Jul 2014
Francis AB Kapur N Hall R
Full Access

Summary Statement

There are no standardised methods for assessing the cement flow behaviour in vertebroplasty. We propose a novel methodology to help understand the interaction of cement properties on the underlying displacement of bone marrow by bone cement in porous media.

Introduction

Concerns related to cement extravasation in vertebroplasty provide the motivation for the development of methodologies for assessing cements (novel and commercially available) and delivery systems. Reproducible and pathologically representative three-dimensional bone surrogates are used to understand the complex rheology underlying the two-phase flow in porous media.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_13 | Pages 18 - 18
1 Mar 2013
Liddle A Borse V Skrzypiec D Timothy J Jacob J Persson C Engqvist H Kapur N Hall R
Full Access

Interbody fusion aims to treat painful disc disease by demobilising the spinal segment through the use of an interbody fusion device (IFD). Diminished contact area at the endplate interface raises the risk of device subsidence, particularly in osteoporosis patients. The aim of the study was to ascertain whether vertebral body (VB) cement augmentation would reduce IFD subsidence following dynamic loading. Twenty-four human two-vertebra motion segments (T6–T11) were implanted with an IFD and distributed into three groups; a control with no cement augmentation; a second with PMMA augmentation; and a third group with calcium phosphate (CP) cement augmentation. Dynamic cyclic compression was applied at 1Hz for 24 hours in a specimen specific manner. Subsidence magnitude was calculated from pre and post-test micro-CT scans. The inferior VB analysis showed significantly increased subsidence in the control group (5.0±3.7mm) over both PMMA (1.6±1.5mm, p=.034) and CP (1.0±1.1mm, p=.010) cohorts. Subsidence in the superior VB to the index level showed no significant differences (control 1.6±3.0mm, PMMA 2.1±1.5mm, CP 2.2±1.2mm, p=.811). In the control group, the majority of subsidence occurred in the lower VB with the upper VB displaying little or no subsidence, which reflects the weaker nature of the superior endplate. Subsidence was significantly reduced in the lower VB when both levels were reinforced regardless of cement type. Both PMMA and CP cement augmentation significantly affected IFD subsidence by increasing VB strength within the motion segment, indicating that this may be a useful method for widening indications for surgical interventions in osteoporotic patients.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_13 | Pages 55 - 55
1 Mar 2013
Skrzypiec D Holub O Liddle A Borse V Timothy J Cook G Kapur N Hall R
Full Access

INTRODUCTION

Over 85% of patients with multiple myeloma (MM) have bone disease, mostly affecting thoraco-lumbar vertebrae. Vertebral fractures can lead to pain and large spinal deformities requiring application of vertebroplasty (PVP). PVP could be enhanced by use of Coblation technique to remove lesions from compromised MM vertebrae prior to cement injection (C-PVP).

METHODS

28 cadaveric MM vertebrae, were initially fractured (IF) up to 75% of its original height on a testing machine, with rate of 1mm/min. Loading point was located at 25% of AP-diameter, from anterior. Two augmentation procedure groups were investigated: PVP and C-PVP. All vertebrae were augmented with 15% of PMMA cement. At the end of each injection the perceived injection force (PIF) was graded on a 5-point scale (1 very easy to 5 almost impossible). Augmented MM vertebrae were re-fractured, following the same protocol as for IF. Failure load (FL) was defined as 0.1% offset evaluated from load displacement curves.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 103 - 103
1 Aug 2012
Hyde P Fisher J Hall R
Full Access

Spinal total disc replacement (TDR) designs rely heavily on total hip replacement (THR) technology and it is therefore prudent to check that typical TDR devices have acceptable friction and torque behaviour. For spherical devices friction factor (f) is used in place of friction coefficient (mju). The range of loading for the lumbar spinal discs is estimated at perhaps 3 times body weight (BW) for normal activity rising to up to 6 times BW for strenuous activity[1]. For walking this equates to around 2000 N, which is the maximum load required by the ISO standard for TDR wear testing[2].

Three Prodisc-L TDR devices (Synthes Spine) were tested in a single station friction simulator. Bovine serum diluted to 25% was used as a lubricating medium. Flexion-extension was ±5 deg for all experiments with constant axial loading of 500, 2000 and 3000 N. The cycle run length was limited to 100 and the f and torque (T) values recorded around the maximum velocity of the cycle point and averaged over multiple cycles.

Preliminary results shows that the 500 N loading produced the largest f of 0.05 ± 0.004. The 2000 N load, which approximates daily activity, gave f = 0.036 ± 0.05 and the 3000 N load gave f = 0.013 ± 0.003. The trend was for lower f with increasing loads.

A lumbar TDR friction factor of 0.036 for a 2000N load and the reduction in f for increasing loads is comparable to the lower end of the range of values reported for THR in similar simulator studies using metal-on-polyethylene bearing materials[3]. The 3000 N result showing that increasing the load above that expected in daily activity does not raise the f could be important when considering rotational stability and anchorage in a TDR device because frictional torque at the bearing surfaces is proportional to the product of load, device radius and f.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVIII | Pages 11 - 11
1 May 2012
Tipper J Vicars R Brown T Ingham E Fisher J Hall R
Full Access

Introduction

The biological response to UHMWPE particles generated by total joint replacements is one of the key causes of osteolysis, which leads to late failure of implants. Particles ranging from 0.1-1.0μm have been shown to be the most biologically active, in terms of osteolytic cytokine release from macrophages [1]. Current designs of lumbar total disc replacements (TDR) contain UHMWPE as a bearing surface and the first reports of osteolysis around TDR in vivo have appeared recently in the literature [2]. The current wear testing standard (ISO18192-1) for TDR specifies only four degrees of freedom (4DOF), i.e. axial load, flexion-extension, lateral bend and axial rotation. However, Callaghan et al. [3] described a fifth DOF, anterior-posterior (AP) shear. The aim of this study was to investigate the effect that this additional AP shear load component had on the size and morphology of the wear particles generated by ProDisc-L TDR devices over five million cycles in a spine simulator.

Methods

A six-station lumbar spine simulator (Simulation Solutions, UK) was used to test ProDisc-L TDR components (Synthes Spine, USA) under the ISO 18192-1 standard inputs and with the addition of an AP load of +175 and −140N. Wear particles were isolated at 2 and 5 mc using a modified alkaline digestion protocol [4]. Particles were collected by filtration and imaged by high resolution FEGSEM. Particle number and volume distributions were calculated as described previously [4] and were compared statistically by one way ANOVA (p<0.05).