Abdul W Goodson M Jones SA
Full Access

Dislocation and instability remain leading cause of failure following THA. We present a single-surgeon 10-year experience with use of Dual Mobility (DM) bearings in Primary and Revision THA using posterior approach.

127 DM bearings were implanted between September 2006 – September 2016; 102 in high-risk primary THA's and 25 revision THA's for either treatment or prevention of instability. Selection for DM bearing followed individual patient risk assessment. Criteria for use of DM bearing were presence of multiple risk factors.

Mean age was 72.9 years. 100 Mono-block DM implants, 22 Modular DM implants and 5 custom-made DM devices were implanted. Revision cohort included those used in conjunction with a cage or porous metal augments. 2 dislocations (1.6%) were observed, both in the Revision group, 1 was recurrent requiring revision to constrained liner. Primary group had 2 revisions; 1 peri-prosthetic fracture and 1 deep infection. No DM bearing specific complications were observed. A constructed life table calculated survival function with endpoint set as revision for any reason demonstrated a cumulative survival of 94% at 7.4 years.

In high-risk patients, DM bearings are successful at preventing and treating dislocation in THA. Primary cohort in this study all had multiple risk factors for instability but no dislocations or bearing specific complications were observed. Dislocations observed in Revision group were associated with major soft tissue deficiency.

This study adds to the promising results already reported with DM THA articulations and should be considered for patients at risk of dislocation or instability.

Runner Up – Best Paper Award