Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_6 | Pages 59 - 59
1 Mar 2017
van Arkel R Ghouse S Ray S Nai K Jeffers J
Full Access

Implant loosening is one of the primary mechanisms of failure for hip, knee, ankle and shoulder arthroplasty. Many established implant fixation surfaces exist to achieve implant stability and fixation. More recently, additive manufacturing technology has offered exciting new possibilities for implant design such as large, open, porous structures that could encourage bony ingrowth into the implant and improve long-term implant fixation. Indeed, many implant manufacturers are exploiting this technology for their latest hip or knee arthroplasty implants. The purpose of this research is to investigate if the design freedoms offered by additive manufacturing could also be used to improve initial implant stability – a precursor to successful long-term fixation. This would enable fixation equivalent to current technology, but with lower profile fixation features, thus being less invasive, bone conserving and easier to revise.

250 cylindrical specimens with different fixation features were built in Ti6Al4V alloy using a Renishaw AM250 additive manufacturing machine, along with 14 specimens with a surface roughness similar to a conventional titanium fixation surface. Pegs were then pushed into interference fit holes in a synthetic bone material using a dual-axis materials testing machine equipped with a load/torque-cell (figure 1). Specimens were then either pulled-out of the bone, or rotated about their cylindrical axis before being pulled out to quantify their ability to influence initial implant stability.

It was found that additively manufactured fixation features could favourably influence push-in/pull-out stability in one of two-ways: firstly the fixation features could be used to increase the amount pull-out force required to remove the peg from the bone. It was found that the optimum fixation feature for maximising pull-out load required a pull-out load of 320 N which was 6× greater than the least optimum design (54 N) and nearly 3× the maximum achieved with the conventional surface (120 N). Secondly, fixation features could also be used to decrease the amount of force required to insert the implant into bone whilst improving fixation (figure 2). Indeed, for some designs the ratio of push-in to pull-out was as high as 2.5, which is a dramatic improvement on current fixation surface technology, which typically achieved a ratio between 0.3–0.6 depending on the level of interference fit. It was also found that the additively manufactured fixation features could influence the level of rotational stability with the optimum design resisting 3× more rotational torque compared to the least optimum design.

It is concluded that additive manufacturing technology could be used to improve initial implant stability either by increasing the anchoring force in bone, or by reducing the force required to insert an implant whilst maintaining a fixed level of fixation. This defines a new set of rules for implant fixation using smaller low profile features, which are required for minimally invasive device design.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_3 | Pages 136 - 136
1 Feb 2017
Ghouse S van Arkel R Babu S Nai K Hooper P Jeffers J
Full Access

Orthopaedic reconstruction procedures to combat osteoarthritis, inflammatory arthritis, metabolic bone disease and other musculoskeletal disorders have increased dramatically, resulting in high demand on the advancement of bone implant technology. In the past, joint replacement operations were commonly performed primarily on elderly patients, in view of the prosthesis survivorship. With the advances in surgical techniques and prosthesis technology, younger patients are undergoing surgeries for both local tissue defects and joint replacements. This patient group is now more active and functionally more demanding after surgery. Today, implanted prostheses need to be more durable (load-bearing), they need to better match the patient's original biomechanics and be able to survive longer.

Additive manufacturing (AM) provides new possibilities to further combat the problem of stress-shielding and promote better bone remodelling/ingrowth and thus long term fixation. This can be accomplished by matching the varying strain response (stiffness) of trabecular or subchondral bone locally at joints. The purpose of this research is therefore to determine whether a porous structure can be produced that can match the required behaviour and properties of trabecular bone regardless of skeletal location and can it be incorporated into a long-term implant.

A stochastic structure visually similar to trabecular bone was designed and optimised for AM (Figure 1) and produced over a range of porosities in multiple materials, Stainless Steel 316, Titanium (Grade 23 – Ti6Al4V ELI) and Commercially Pure Titanium (Grade 2) using a Renishaw AM250 metal additive manufacturing system. Over 150 cylindrical specimens were produced per material and subjected to a compression test to determine the specimens' Elastic Modulus (Stiffness) and Compressive Yield Strength. Micro-CT scans and gravimetric analysis were also performed to determine and validate the specimens' porosity. Results were then graphed on a Strength vs. Stiffness Ashby plot (Figure 2) comparing the values to those of trabecular bone in the tibia and femur.

It was found that AM can produce porous structures with an elastic modulus as low as 100 MPa up to 2.7 GPa (the highest stiffness investigated in this study). Titanium structures with a stiffness <500MPa had compressive strengths towards the bottom range of similar stiffness trabecular bone. Between 500 MPa − 1 GPa Titanium AM porous structures match the compressive strength of equivalent stiffness trabecular bone and from 1 GPa − 2 GPa the Ti structures exceed the strength of equivalent stiffness trabecular bone up to ∼2.5 times and consequently increase by a power law.

These results show that AM can produce structures with similar stiffness to trabecular bone over a range of skeletal locations whilst matching or exceeding the compressive strength of bone. The results have not yet taken into account fatigue life with the fatigue life of these types of structures tending to be between 0.1 – 0.4 of their compressive strength. This means that a titanium porous structure would need to be 2.5 – 10 times stiffer or stronger than the portion of trabecular bone it is replacing. This data is highly encouraging for AM manufactured, bone stiffness matched implant technology.