The purpose of this study was to determine long-term survival free from all-cause revision and stem-related failure, as well as radiographic and functional outcomes of the ZMR stem in revision THA. We retrospectively identified all patients in our institutional database who underwent revision THA using the ZMR Revision Hip system from the year 2000 to 2007 (minimum two-year follow-up). Of the 112 ZMR hips (110 patients) identified, a total of 106 hips (104 patients) met inclusion criteria. The mean study follow-up 13.9 years (range 2–22). Indications for index ZMR revision included aseptic loosening (72.1%), periprosthetic joint infection (17.3%), periprosthetic fracture (9.6%), and stem fracture (1.0%). Kaplan-Meier analysis was used to determine the all-cause and stem-related failure revision-free survival. Validated patient reported outcomes were collected and available radiographs were reviewed to determine implant stability. A total of 17 hips (16.0%) underwent a re-revision of any component. The indications for re-revision were stem failure (35.3%), infection (29.4%), instability (29.4%), and aseptic loosening of the acetabular component (5.9%). The five- and 15-year all-cause survivorship was 89.5% (95% CI 86.5–92.5) and 83.0% (95% CI 79.2–86.8), respectively. There were six re-revisions for stem failure (5.7%); five for stem fracture and one for aseptic loosening. The average time to stem failure was 4.6 years (range 1.3–8.2). The five- and 15-year survivorship free from stem-related failure was 97.2% (95% CI 95.6–98.8) and 93.9% (95% CI 91.5–96.3), respectively. At final follow-up the mean Oxford hip score was 36.9 and all surviving ZMR stems were well fixed on radiographs. Femoral revision with the ZMR stem offers satisfactory long-term survivorship and promising clinical outcomes. Although uncommon, stem fracture was the most common reason for stem-related failure.
Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high early failure rates continues to undermine confidence in use. A large contributor to failure is adverse local tissue reactions around metal-on-metal (MoM) bearing surfaces. Such phenomena have been well-explored around MoM total hip arthroplasties, but comparable data in equivalent hip resurfacing procedures is lacking. In order to define genetic predisposition, we performed a case-control study investigating the role of human leucocyte antigen (HLA) genotype in the development of pseudotumours around MoM hip resurfacings. A matched case-control study was performed using the prospectively-collected database at the host institution. In all, 16 MoM hip resurfacing 'cases' were identified as having symptomatic periprosthetic pseudotumours on preoperative metal artefact reduction sequence (MARS) MRI, and were subsequently histologically confirmed as high-grade aseptic lymphocyte-dominated vasculitis-associated lesions (ALVALs) at revision surgery. ‘Controls’ were matched by implant type in the absence of evidence of pseudotumour. Blood samples from all cases and controls were collected prospectively for high resolution genetic a nalysis targeting 11 separate HLA loci. Statistical significance was set at 0.10 a priori to determine the association between HLA genotype and pseudotumour formation, given the small sample size.Aims
Methods
Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up. We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.Aims
Methods
The practice of overlapping surgery has been increasing in the delivery of orthopaedic surgery, aiming to provide efficient, high-quality care. However, there are concerns about the safety of this practice. The purpose of this study was to examine the safety and efficacy of a model of partially overlapping surgery that we termed ‘swing room’ in the practice of primary total hip (THA) and knee arthroplasty (TKA). A retrospective review of prospectively collected data was carried out on patients who underwent primary THA and TKA between 2006 and 2017 in two academic centres. Cases were stratified as partially overlapping (swing room), in which the surgeon is in one operating room (OR) while the next patient is being prepared in another, or nonoverlapping surgery. The demographic details of the patients which were collected included operating time, length of stay (LOS), postoperative complications within six weeks of the procedure, unplanned hospital readmissions, and unplanned reoperations. Fisher's exact, Wilcoxon rank-sum tests, chi-squared tests, and logistic regression analysis were used for statistical analysis.Aims
Methods
The practice of overlapping surgery has been increasing in the delivery of orthopaedic care, aiming to provide efficient, high-quality care. However, there have been concerns about the safety of this practice. The purpose of this study is to examine safety and efficacy of a model of partially overlapping surgery that we termed “the swing room” in practice in primary hip and knee arthroplasty. A retrospective review of prospectively collected data using an administrative database was carried out on patients who underwent primary unilateral total hip and total knee arthroplasty from 2006 to 2017 at two sites of one academic center staffed by four arthroplasty surgeons. All revisions and bilateral primary procedures were excluded. Cases were stratified as overlapping or non-overlapping. Overlapping was defined when a surgeon had access to two operating rooms with two teams, and non-overlapping was defined as when a surgeon only had access to a single operating room on a particular day. Patient demographic characteristics, operating room time, procedure time, length of stay, Postoperative complications within 30 days of the procedure, unplanned hospital readmissions, unplanned reoperations, and emergency department visits were collected. The Fisher's exact Wilcoxon rank-sum test and logistic regression analysis were used for statistical analysis.Introduction
Methods
We investigated the long-term performance of the Tripolar Trident acetabular component used for recurrent dislocation in revision total hip arthroplasty. We assessed: 1) rate of re-dislocation; 2) incidence of complications requiring re-operation; and 3) Western Ontario and McMaster Universities osteoarthritis index (WOMAC) pain and functional scores. We retrospectively identified 111 patients who had 113 revision tripolar constrained liners between 1994 and 2008. All patients had undergone revision hip arthroplasty before the constrained liner was used: 13 after the first revision, 17 after the second, 38 after the third, and 45 after more than three revisions. A total of 75 hips (73 patients) were treated with Tripolar liners due to recurrent instability with abductor deficiency, In addition, six patients had associated cerebral palsy, four had poliomyelitis, two had multiple sclerosis, two had spina bifida, two had spondyloepiphyseal dysplasia, one had previous reversal of an arthrodesis, and 21 had proximal femoral replacements. The mean age of patients at time of Tripolar insertions was 72 years (53 to 89); there were 69 female patients (two bilateral) and 42 male patients. All patients were followed up for a mean of 15 years (10 to 24). Overall, 55 patients (57 hips) died between April 2011 and February 2018, at a mean of 167 months (122 to 217) following their tripolar liner implantation. We extracted demographics, implant data, rate of dislocations, and incidence of other complications.Aims
Patients and Methods
Fully constrained liners are used to treat recurrent dislocations or patients at high risk after total hip replacements. However, they can cause significant morbidities including recurrent dislocations, infections, aseptic loosening and fractures. We examine long term results of 111 patients with tripolar constrained components to assess their redislocation and failure rate. The purpose of this study was to assess survivorship, complications and functional outcomes at a minimum 10 years after the constrained tripolar liners used in our institute.Background
Questions/purposes
High complication rates and poor outcomes have been widely reported in patients undergoing revision of large head metal-on-metal arthroplasty. A previous study from our center showed high rates of dislocation, nerve injury, early cup loosening and pseudotumor recurrence. After noting these issues, we implemented the following changes in surgical protocol in all large head MOM revisions: 1. Use of highly porous shells in all cases 2. Use of largest femoral head possible 3. Low threshold for use of dual mobility and constrained liners when abductors affected or absent posterior capsule 4. Use of ceramic head with titanium sleeve in all cases 5. Partial resection of pseudotumor adjacent to sciatic and femoral nerves. The purpose of the present study is to compare the new surgical protocol above to our previously reported early complications in this group of patients We specifically looked at (1) complications including reoperations; (2) radiologic outcomes; and (3) functional outcomes. Complication rates after (Group 1), and before (Group 2) modified surgical protocol were compared using Chi-square test, assuming statistical significance p<0.05.Background
Questions/purposes
Fretting and corrosion at the modular head/neck junction, known
as trunnionosis, in total hip arthroplasty (THA) is a cause of adverse
reaction to metal debris (ARMD). We describe the outcome of revision
of metal-on-polyethylene (MoP) THA for ARMD due to trunnionosis
with emphasis on the risk of major complications. A total of 36 patients with a MoP THA who underwent revision
for ARMD due to trunnionosis were identified. Three were excluded
as their revision had been to another metal head. The remaining
33 were revised to a ceramic head with a titanium sleeve. We describe
the presentation, revision findings, and risk of complications in
these patients.Aims
Patients and Methods
Reconstruction of the acetabulum after failed total hip arthroplasty
(THA) can be a surgical challenge in the presence of severe bone
loss. We report the long-term survival of a porous tantalum revision
acetabular component, its radiological appearance and quality of
life outcomes. We reviewed the results of 46 patients who had undergone revision
of a failed acetabular component with a Paprosky II or III bone
defect and reconstruction with a hemispherical, tantalum acetabular
component, supplementary screws and a cemented polyethylene liner.Aims
Patients and Methods
We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods
Tapered fluted titanium stems are increasingly
used for femoral revision arthroplasty. They are available in modular and
non-modular forms. Modularity has advantages when the bone loss
is severe, the proximal femur is mis shapen or the surgeon is unfamiliar
with the implant, but it introduces the risk of fracture of the
stem at the junction between it and the proximal body segment. For
that reason, and while awaiting intermediate-term results of more recently
introduced designs of this junction, non-modularity has attracted
attention, at least for straightforward revision cases. We review the risks and causes of fracture of tapered titanium
modular revision stems and present an argument in favour of the
more selective use of modular designs. Cite this article:
Adverse reaction to wear and corrosion debris
is a cause for concern in total hip arthroplasty (THA). Modular junctions
are a potential source of such wear products and are associated
with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our
unit for this complication following metal-on-highly cross-linked
polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as
infection, and present the aggregate laboratory results and pathological
findings in this series. The clinical presentation was pain, swelling or instability.
Solid, cystic and mixed soft-tissue lesions were noted on imaging
and confirmed intra-operatively. Corrosion at the head–neck junction
was noted in all cases. No bacteria were isolated on multiple pre-
and intra-operative samples yet the mean erythrocyte sedimentation
rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and
stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft–tissue reactions can occur in MoP THA owing to corrosion
products released from the head–neck junction. The diagnosis should
be carefully considered when investigating pain after THA. This
may avoid the misdiagnosis of periprosthetic infection with an unidentified
organism and mitigate the unnecessary management of these cases
with complete single- or two-stage exchange. Cite this article:
Tapered, fluted, modular, titanium stems are
increasingly popular in the operative management of Vancouver B2 and
selected B3 peri-prosthetic femoral fractures. We have reviewed
the results at our institution looking at stem survival and clinical
outcomes and compared this with reported outcomes in the literature.
Stem survival at a mean of 54 months was 96% in our series and 97%
for combined published cases. Review of radiology showed maintenance
or improvement of bone stock in 89% of cases with high rates of
femoral union. Favourable clinical outcome scores have reported
by several authors. No difference in survival or clinical scores
was observed between B2 and B3 fractures. Tapered stems are a useful
option in revision for femoral fracture across the spectrum of femoral
bone deficiency. Cite this article:
Tapered, fluted, modular, titanium stems have
a long history in Europe and are increasing in popularity in North America.
We have reviewed the results at our institution looking at stem
survival and clinical outcomes. Radiological outcomes and quality
of life assessments have been performed and compared to cylindrical
non-modular cobalt chromium stems. Survival at five years was 94%.
This fell to 85% at ten years due to stem breakage with older designs.
Review of radiology showed maintenance or improvement of bone stock
in 87% of cases. Outcome scores were superior in tapered stems despite
worse pre-operative femoral deficiency. Tapered stems have proved
to be a useful alternative in revision total hip arthroplasty across
the spectrum of femoral bone deficiency.
We report the outcome at ten to 15 years of two-stage revision for hip infection in 99 patients using the Prostalac articulated hip spacer system. All the patients were contacted to determine their current functional and infection status using the Oxford-12, Short form-12, and Western Ontario and McMaster University Osteoarthritis Index questionnaires. A total of 11 of the 99 patients had a further infection, of whom seven responded to repeat surgery with no further sequelae. The mean interval between the stages was five months (1 to 36). We were able to review 48 living patients, with a mean age of 72 years (46 to 86), 34 (71%) of whom provided health-related quality-of-life outcome scores. The mean follow-up was 12 years (10 to 15). The long-term success rate was 89% and with additional surgery this rose to 96%. The mean global Western Ontario and McMaster University Osteoarthritis Index score was 80.6 ( Two-stage revision for hip infection using a Prostalac interim spacer offers a predictable and lasting solution for patients with this difficult problem.
A prospective cohort of 222 patients who underwent revision hip replacement between April 2001 and March 2004 was evaluated to determine predictors of function, pain and activity level between one and two years post-operatively, and to define quality of life outcomes using validated patient reported outcome tools. Predictive models were developed and proportional odds regression analyses were performed to identify factors that predict quality of life outcomes at one and two years post-operatively. The dependent outcome variables were the Western Ontario and McMaster Osteoarthritis Index (WOMAC) function and pain scores, and University of California Los Angeles activity scores. The independent variables included patient demographics, operative factors, and objective quality of life parameters, including pre-operative WOMAC, and the Short Form-12 mental component score. There was a significant improvement ( Predictors of quality of life outcomes after revision hip replacement were established. Although some patient-specific and surgery-specific variables were important, age, gender, Charnley class and pre-operative WOMAC function score had the most robust associations with outcome.
Removal of well-fixed, cementless, acetabular components during revision arthroplasty remains a challenging problem. Further damage to host bone may limit options for reconstruction and compromise the long-term result of the revision operation. We report the results of 31 hips with well-fixed, cementless sockets which were removed using a new cup extraction system. In all hips the socket was removed without difficulty and with minimal further bone loss.
We have undertaken a prospective evaluation of health-related quality-of-life outcomes in 26 consecutive patients undergoing periacetabular osteotomy with a minimum follow-up of 24 months (24 to 42). The Western Ontario-Macmaster (WOMAC) and Short Form 36 (SF-36) outcome measures were evaluated preoperatively and at follow-up. Secondary outcome measures included retrospective assessment of sports activity and satisfaction with the procedure. Follow-up data were available for 22 patients (85%). A significant improvement in the pain (p <
0.0001) and function (p <
0.0001) scales of the WOMAC and the SF-36 physical score (p <
0.0001) was seen. The mean sports activity score improved from 1.9 to 4.4, with a very high mean satisfaction rating of 89.7/100. A significant improvement in both global and disease-specific health outcomes can be expected after this operation, with a high rate of satisfaction and improvement in sporting activities.
There are few medium- and long-term data on the outcome of the use of proximal femoral structural allografts in revision hip arthroplasty. This is a study of a consecutive series of 40 proximal femoral allografts performed for failed total hip replacements using the same technique with a minimum follow-up of five years (mean 8.8 years; range 5 to 11.5 years). In all cases the stem was cemented into both the allograft and the host femur. The proximal femur of the host was resected in 37 cases. There were four early revisions (10%), two for infection, one for nonunion of the allograft-host junction, and one for allograft resorption noted at the time of revision of a failed acetabular reconstruction. Junctional nonunion was seen in three patients (8%), two of whom were managed successfully by bone grafting, and bone grafting and plating respectively. Instability was observed in four (10%). Trochanteric nonunion was seen in 18 patients (46%) and trochanteric escape in ten of these (27%). The mean Harris hip score improved from 39 to 79. Severe resorption involving the full thickness of the allograft was seen in seven patients (17.5%). This progressed rapidly and silently, but has yet to cause failure of any of the reconstructions. Profound resorption of the allograft may be related to a combination of factors, including a slow form of immune rejection, stress shielding and resorption due to mechanical disuse with solid cemented distal fixation, and the absence of any masking or protective effect which may be provided by the retention of the bivalved host bone as a vascularised onlay autograft. Although continued surveillance is warranted, the very good medium-term clinical results justify the continued use of structural allografts for failed total hip replacements with severe loss of proximal femoral bone.