Parchi P Andreani L Cutolo F Carbone M Ferrari V Ferrari M Lisanti M
Full Access

Aim of the study was the evaluation of the efficacy of the use of a new wearable AR video see-throught system based on Head Mounted Displays (HMDs) to guide the position of a working cannula into the vertebral body through a transpedicular approach without the use X-Ray images guidance.

We describe a head mounted stereoscopic video see-through display that allows the augmentation of video frames acquired by two cameras with the rendering of patient specific 3D models obtained on the basis of pre-operative radiological volumetric images. The system does not employ any external tracker to detect movements of the user or of the patient. User's head movements and the consistent alignment of the virtual patient with the real one, are accomplished through machine vision methods applied on pairs of live images.

Our system has been tested on an experimental setup that simulate the reaching of lumbar pedicle as in a vertebral augmentation procedure avoiding the employment of ionizing radiation. Aim of the study is to evaluate the ergonomics and the accurancy of the systems to guide the procedure. We performed 4 test sessions with a total of 32 kirschner wire implanted by a single operator wearing the HMD with the AR guide. The system accurancy was evaluated by a post-operative CT scan.

The most ergonomic AR visualization comprise the use of a pair of virtual viewfinders (one at the level of the skin entry point and one at the level of the trocar's bottom) aligned according to the planned direction of the trocar insertion. With such AR guide the surgeon must align the tip of the needle to the center of the first viewfinder placed on the patient's skin. indeed the viewfinder barycenter provides a 2 degrees of freedom (DoFs) positioning guide corresponding to the point of insertion preoperatively planned over the external surface of the model. The second viewfinder is used by the surgeon to rotate and align the trocar according to the planned direction of insertion (2 rotational DOFs). After the first test series a clamping arm has been introduced to maintain the reached trocar's trajectory.

The post-operative CT scan was registered to the preoperative one and the trajectories obtained with the AR guide were compared to the planned one. The overal results obtained in the 4 test session show a medium error of 1.18+/−0.16 mm.

In the last year there was a growing interest to the use of Augmented Reality systems in which the real scene watched by the surgeon is merged with virtual informations extracted from the patient's medical dataset (medical data, patient anatomy, preoperative plannig). Wearable Augmented Reality (WAR) with the use of HDMs allows the surgeon to have a “natural point of view” of the surgical field and of the patient's anatomy avoiding the problems related to eye-hand coordination.

Results of the in vitro tests are encouraging in terms of precision, system usability and ergonomics proving our system to be worthy of more extensive tests.


Orthopaedic Proceedings
Vol. 97-B, Issue SUPP_15 | Pages 4 - 4
1 Dec 2015
Ferrari M Astore F Santoro G Eusebio A Meda F Grappiolo G
Full Access

Almost any kind of bacteria can be involved in prosthetic joint infections (PJI).

In 2014 we collected data on 3 patients presenting with hip or knee pain and radiological signs of prosthetic joint mobilization without any other clinical or biochemical suspect of infection.

Interestingly, in all cases we isolated bacteria very rarely involved in human infections: Cupriavidus pauculus, Ralstonia paucula and Neisseria flavescens. Both Cupriavidus pauculus and Ralstonia paucula are Gram negative environmental bacteria which can contaminate water (tap, pool and mineral) and have been isolated from a variety of human clinical sources including blood, wounds, sputum, urine, eye, throat and peritoneal fluid, as non pathogenic flora. In such two patients we were able to cure the infection with quinolones (both strains were sensible). Neisseria species are Gram-negative bacteria included among the proteobacteria. Neisseria flavescens is often found in the upper respiratory tract and oropharynx of humans, rarely associated with infectious process (necrotizing pneumonia and empyema). In this last patient we successfully used piperacillin-tazobactam i.v. during hospitalization and shifted to amoxicillin-clavulanate per os at discharge.

In conclusion, any bacteria can induce PJI: in some cases pre-operative blood tests are normal and therefore useless. We need new tests to overcome this diagnostic problem (such as alpha-defensin) and we must check on rare bacteria prolonging incubation to a more extended period of time (as in Propionebacterium acnes or Mycobacterium spp) and/or sending samples to specialized laboratories.

We thank all the medical and nurse staff of Prosthetic Joint Replacement Unit of Orthopaedic Center, Humanitas Research Hospital


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLIV | Pages 20 - 20
1 Oct 2012
Ferrari V Parchi P Condino S Carbone M Baluganti A Ferrari M Mosca F Lisanti M
Full Access

Pedicle screws fixation to stabilise lumbar spinal fusion is the gold standard for posterior stabilisation. Pedicle screws are today positioned in free hand or under fluoroscopic guidance with an error from 20% up to 40–50%, which can determine the inefficacy of treatment or severe damages to close neurologic structures. Surgical navigation drastically increases screws placement accuracy. However its clinical application is limited due to cost reasons and troubles related to the presence of a localiser in the OR and the need to perform a registration procedure before surgery. An alternative image guided approach is the use of patient specific templates similar to the ones used for dental implants or knee prosthesis. Until now, the proposed solutions allow to guide the drill, and in some cases, as templates fit completely around vertebra, they require the complete removal of soft tissues on a large portion of the spine, so increasing intervention invasiveness. To reduce the soft tissue demolition, some authors proposed a fitting based on small “V shape” contact points, but these solutions can determine instability of the template and the reacting of wrong stable positions.

In our solution, after spine CT acquisition, each vertebra is segmented using a modified version of ITK-SNAP software, on which the surgeon plans screws positioning and finally the template is designed around the chosen trajectories, using a tool which allows to insert cylinders (full or empty) in the segmented images. Each template, printed in ABS, contains two hollow cylinders, to guide the screws, and multiple contact points on the bone surface, for template stabilisation.

We made an in-vitro evaluation on synthetic spine models (by Sawbones) to study different template designs. During this first step an ongoing redesign allowed to obtain an optimal template stability and an easy template positioning to minimise the intervention invasiveness. A first contact point, which fits on the sides of the spinous process, is used to simplify template alignment. The other 4 contact points, which consists of cylinders (diameter 5 mm), fit exactly on spine surface in correspondence to the vertebra's lamina and articular processes to stabilise the template in an unique position. Templates can be used to guide not only the drill, but also Kirschner wires, to guide cannulated screws. After the Kirschner wires insertion the template can be dismounted for its removal (the direction of the kirschner wires are not parallel).

After the definitive template design an ex-vivo animal test on 2 porcine specimens has been conducted to evaluate template performance in presence of soft-tissue in place. The specimens have been scanned with CT, we realised a total of 14 templates and we performed the insertion of 28 Kirschner wires. We evaluated that after the soft tissue dissection and the bone exposure, the template can be easily positioned in the right unique position, with no additional tissue removal compared to the traditional approach, requiring just removal of the soft tissue under the small contact points using an electric cutter. The surgeon evaluated (and corrected) some wrong stable template positions when not all the contact points were in contact with the bone surface. The post-op evaluation was made with a CT scan that showed 1 cortical pedicle violation (3.5%) (grade II according to the FU classification).