Variation in pelvic tilt during postural changes may affect functional alignment. The primary objective of this study was to quantify the changes in lumbo-pelvic-femoral alignment from sitting to standing in patients undergoing THA. 144 patients were enrolled. Standing and sitting radiographs using the EOS imaging system were analyzed preoperatively and 1-year postoperatively. Pelvic incidence (PI), lumbar lordosis (LL), sacral slope (SS), proximal femoral angle (PFA) and spine/femoroacetabular flexion were determined. 38 patients had multilevel DDD (26%). Following THA, patients sat with increased anterior pelvic tilt demonstrated by a significant increase in sitting lumbar lordosis (28° preop vs 35° postop; p<0.01) and sacral slope (18° vs 23°; p<0.01). Following THA, patients flexed less through their spines (preop 26° vs postop 19°; p<0.01) and more through their hips (femoroacetabular flexion) (preop 60° vs postop 67°; p<0.01) to achieve sitting position. Patients with multilevel DDD sat with less spine flexion (normal 22° vs spine 13°; p<0.01), less change in sacral slope (more relative anterior tilt) (17° vs 9°; p<0.01), and more femoroacetabular flexion (64° vs 71°; p<0.01). For the majority of patients after THA, a larger proportion of lumbo-pelvic-femoral flexion necessary to achieve a sitting position is derived from femoroacetabular flexion with an associated increase in anterior pelvic tilt and a decrease in lumbar spine flexion. These changes are more pronounced among patients with multilevel DDD. Surgeons may consider orienting the acetabular component with greater anteversion and inclination in patients identified preoperatively to have anterior pelvic tilt or significant DDD.
We report a case of a male patient presenting
with bilateral painful but apparently well-positioned and -fixed
large-diameter metal-on-metal hip replacements four years post-operatively.
Multiple imaging modes revealed a thick-walled, cystic expansile
mass in communication with the hip joint (a pseudotumour). Implant
retrieval analysis and tissue culture eliminated high bearing wear
or infection as causes for the soft-tissue reaction, but noted marked corrosion
of the modular neck taper adaptor and corrosion products in the
tissues. Therefore, we believe corrosion products from the taper
caused by mismatch of the implant components led to pseudotumour
formation requiring revision.