Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_14 | Pages 90 - 90
1 Dec 2019
Langvatn H Schrama JC Engesæter LB Hallan G Furnes O Lingaas E Walenkamp G Dale H
Full Access

Aim

The aim of this study was to assess the influence of the true operating room (OR) ventilation on the risk of revision due to infection after primary total hip arthroplasty (THA) reported to the Norwegian Arthroplasty Register (NAR).

Method

40 orthopedic units were included during the period 2005 – 2015. The Unidirectional airflow (UDAF) systems were subdivided into small-area, low-volume, vertical UDAF (lvUDAF) (volume flow rate (VFR) (m3/hour) <=10,000 and diffuser array size (DAS) (m2) <=10); large-area, high-volume, vertical UDAF (hvUDAF) (VFR >=10,000 and DAS >=10) and Horizontal UDAF (H-UDAF). The systems were compared to conventional, turbulent ventilation (CV) systems. The association between revision due to infection and OR ventilation was assessed using Cox regression models, with adjustments for sex, age, indication for surgery, ASA-classification, method of fixation, modularity of the components, duration of surgery, in addition to year of primary THA. All included THAs received systemic, antibiotic prophylaxis.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_23 | Pages 50 - 50
1 Dec 2016
Lutro O Dale H Sjursen H Schrama JC Høvding P Bartz-Johannessen CA Hallan G Engesæter LB
Full Access

Aim

To see what surgical strategy was used in treating infected total hip arthroplasties (THA), relative to bacterial findings, level of inflammation, length of antibiotic treatment (AB) and re-revisions. Further, to assess the results of treatment after three months and one year.

Method

We used our national arthroplasty register (NAR) to identify THA revised for deep infection from 2004–2015 reported from our hospital. We identified the strategy of revision, i.e. one-stage exchange (one-stage), two-stage exchange (two-stage), debridement and implant retention (DAIR), or Girdlestone, and reported re-revisions for infection. We defined cure as no AB, no need for further surgery and joint with prosthesis (not Girdlestone).

From the hospitals’ medical records we retrieved bacterial findings from the revisions, level of C-reactive protein (CRP), type of antibiotics given, duration of antibiotic therapy and clinical data regarding the patients. The information reported to the NAR was also validated.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_23 | Pages 87 - 87
1 Dec 2016
Langvatn H Schrama JC Engesæter LB Lingaas E Dale H
Full Access

Aim

The aim of this study was to validate the information on operating room ventilation reported to the Norwegian Arthroplasty Register (NAR) and to assess the influence of this ventilation on the risk of revision due to infection after primary total hip arthroplasty (THA).

Method

Current and previous ventilation systems were evaluated together with the hospitals head engineer in 40 orthopaedic hospitals. The ventilation system of each operating room was assessed and confirmed as either conventional ventilation, vertical laminar airflow (LAF) or horizontal LAF. We then identified cases of first revision due to deep infection after primary THA and the type of ventilation system reported to the NAR in the period 1987–2014. The association between revision due to infection and operating room ventilation was estimated by relative risks (RR) in a Cox regression model.