Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_6 | Pages 61 - 61
1 Apr 2018
Upmann C Eisele M Pandorf T
Full Access

Introduction

Ceramic ball heads are well known in hip arthroplasty for their superior tribology performance and high burst strength. To assess the ball head performance and the in-vivo fracture risk Pandorf et al 2008 examined the burst strength of BIOLOX®forte (pure aluminium oxide ceramic, CeramTec GmbH, Plochingen, Germany) ball heads on clean standard test tapers and contaminated test tapers. They found that fat tissue and scratches are reducing the burst strength to 40% and to 20% of the reference burst strength, respectively. The aim of this work is to investigate if BIOLOX®delta (alumina matrix ceramic, CeramTec GmbH, Plochingen, Germany) ball heads show a similar behaviour as BIOLOX®forte ball heads with respect to taper contamination.

Materials and Methods

Each test series consisted of n=5 BIOLOX®delta 28–12/14 L ball heads and n=5 metal test tapers (Ti-6Al-4V, ISO 5832-3). For the reference series the metal tapers remained untouched representing the CeramTec standard test procedure. For the fluid series the ball heads were filled up with tap water or calf blood serum. For the solid series the metal test tapers were contaminated with small particles of bovine bone, commercially available bone cement and porcine fat tissue in the engagement zone. A chisel and a slight hammer tap were used to scratch the proximal region of the metal test taper. The ball heads were then manually attached to the contaminated metal test tapers without further force appliances. An apparatus according to ISO 7206-10 was used for burst testing. The tests were performed at CeramTec in-house test laboratory.