Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_4 | Pages 134 - 134
1 Jan 2016
Wimmer M Freed R Daniels C Pourzal R
Full Access

Introduction

Current pre-clinical testing is performed using knee wear simulators with standardized walking profiles. Differences in generated damage patterns to those observed on retrieved liners have been explained with the absence of activities other than walking, less severe loading conditions, and a discrepancy in the simulator's tibiofemoral contact mechanics and in vivo knee excursion. While it has been recognized that rotational alignment of the knee may also drive the location and shape of wear scars, to the best of our knowledge this parameter has not been investigated in knee simulator studies.

Methods

Here, we use patient specific gait as input to the simulation to approximate the patient specific contact mechanics. Kinematic and kinetic input data was obtained from gait analysis of a patient with a MGII (Zimmer Inc.) prosthesis at 11 years post-op using the point cluster technique for tibiofemoral kinematics, and a mathematical model for internal force calculations. Using the identical type of prosthesis on the simulator, wear tests were conducted in displacement mode on a closed-loop controlled station. Because x-rays of the patient suggested an internal rotation of the tibial tray, it was varied form 0–10° and the effect on location and wear scar dimension was assessed. Results were compared with the retrieved liner (obtained after 13 years in vivo).