Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_28 | Pages 17 - 17
1 Aug 2013
Ferretti A Iorio R Mazza D Caperna L Bolle G Argento G Conteduca F
Full Access

Introduction

The aim of this study is to investigate the accuracy and reliability of a Magnetic Resonance Imaging (MRI) based Patient Match Technology (PMT) system (VISIONAIRE, Smith & Nephew, Inc, Memphis, Tenn) by intraoperative use of VectorVision knee navigation software from BrainLAB (Redwood City, California, USA).

Methods

Between February 2011 and May 2011, 15 patients with primary gonarthrosis were selected for unilateral Total Knee Arthroplasty (TKA). The first three patients were excluded from this study, as they were considered as a warm up to set up the procedure. Therefore 12 patients entered the study. Preoperatively all patient underwent a full-length weight-bearing radiograph in antero-posterior (AP) and a MRI according to the protocol suggested and approved by the manufacturer. All patients were operated with cemented posterior stabilised prosthesis cruciate ligament sacrificing (Journey BCS, Smith & Nephew, Inc, Memphis, Tennessee, USA) by the same surgeon using the VISIONAIRE patient matched cutting jigs. During surgery, once the guides were placed and fixed, the orientation was checked by the navigator. The following parameters were evaluated: size of the implant, alignment in coronal and sagittal plane. An unsatisfactory result was considered an error ≥ 2° in both plane for each component as a possible error of 4° could result in aggregate.